SonTek/YSI
ADVField/Hydra Operation Manual (September 1, 2001)
77
Each current meter purchased from SonTek that has the
DRUCK
sensor installed already has the
appropriate calibration coefficients loaded into the processor. If this information is lost, you can
use the calibration sheet provided by the pressure sensor vendor to construct your own calibration
file. The structure of the
DRUCK
calibration file, required by SonTek systems, is as follows.
•
Pressure Sensor Serial Number
•
Oscillator Frequency (in Hz)
•
X
•
Y
•
Low end of temperature range
•
Corresponding voltage of the diode
•
Diode sensitivity
•
K00 K01 K02 K03
•
K10 K11 K12 K13
•
K20 K21 K22 K23
•
K30 K31 K32 K33
•
K40 K41 K42 K43
•
K50 K51 K52 K53
Example – Druck calibration file
1145118.drk
:
1145118
12.006
+8.803
+5.502
-10
622
-2.21
+1.603
-2.0699499E-02
+5.5220000E-05
-5.6160001E-08
+1.800
-5.8612000E-06
+3.6197001E-08
-3.8715001E-10
+3.8615943E-04
-1.5274000E-09
-4.7059999E-12
-1.3233001E-13
+7.0553313E-08
-1.9064000E-12
-4.0058001E-14
+9.0119996E-16
+1.7913088E-11
-1.3525000E-15
+1.7647000E-17
+3.8552999E-19
+3.7756621E-15
+1.9573000E-19
+3.6951001E-20
-3.3300999E-22
4.9.4. Paroscientific Digiquartz Ext. Freq. Pressure Sensor (PAROSFREQ)
User requirements for deeper ratings (compared to
DRUCK
) and faster response (compared to
PAROS
) convinced SonTek to offer the Paroscientific digiquartz sensor with frequency output
(
PAROSFREQ
). Functionally, the interface for the
PAROSFREQ
is identical to the interface for the
DRUCK
, as both are frequency-based sensors. The
PAROSFREQ
sensors are available for ADV and
Hydra systems. The differences, when compared with a strain-gage sensor, are:
•
The
PAROSFREQ
is an absolute sensor. It reads 1 dbar in air; a strain gage reports 0 dbars.
•
Requires a special interface to sample the frequency signal.
•
Requires a special frequency splitter cable.
•
Sampling rate is limited to 10 Hz (ADVs and Hydras) because of the time required to sample
the frequency signal with sufficient precision.
•
Requires a calibration (specific to each sensor) to be loaded into the processor (system CPU
or a user PC) to obtain valid pressure readings.
•
Available in 50-psia (
≈
22-m sealed gage) and 100-psia (
≈
60-m sealed gage) pressure ratings.
Note: The X, Y, and K coefficients can
be found on the Druck calibration sheet.
Содержание SonTek ADVField
Страница 2: ......
Страница 8: ...SonTek YSI iv ADVField Hydra Operation Manual September 1 2001 NOTES ...
Страница 12: ......
Страница 32: ...SonTek YSI 20 ADVField Operation Manual September 1 2001 ...
Страница 102: ...SonTek YSI 90 ADVField Operation Manual September 1 2001 ...
Страница 104: ...SonTek YSI 92 ADVField Operation Manual September 1 2001 ...
Страница 118: ...SonTek YSI 106 ADVField Operation Manual September 1 2001 ...
Страница 122: ...SonTek YSI 110 ADVField Operation Manual September 1 2001 ...
Страница 126: ...SonTek YSI 2 ADVField Operation Manual September 1 2001 ...
Страница 132: ...SonTek YSI vi ADVField Software Manual September 1 2001 NOTES ...
Страница 136: ...SonTek YSI ADVField Software Manual September 1 2001 4 ...
Страница 152: ...SonTek YSI ADVField Software Manual September 1 2001 20 ...
Страница 160: ...SonTek YSI ADVField Software Manual September 1 2001 28 ...
Страница 162: ...SonTek YSI ADVField Software Manual September 1 2001 30 ...
Страница 164: ...SonTek YSI ADVField Software Manual September 1 2001 32 ...
Страница 168: ...SonTek YSI ADVField Software Manual September 1 2001 36 ...
Страница 174: ...SonTek YSI ADVField Software Manual September 1 2001 42 ...
Страница 182: ...SonTek YSI ADVField Software Manual September 1 2001 50 ...
Страница 184: ...SonTek YSI ADVField Software Manual September 1 2001 52 Figure 15 ViewHydra Screen Shot ...
Страница 196: ...SonTek YSI ADVField Software Manual September 1 2001 64 ...
Страница 198: ...SonTek YSI ADVField Software Manual September 1 2001 66 ...
Страница 202: ......
Страница 206: ......
Страница 220: ...SonTek YSI Acoustic Doppler Velocimeter Principles of Operation September 1 2001 14 ...