4.7.
Optional Compass/Tilt Sensor for the ADVField
The ADVField can be equipped with a compass/tilt sensor to translate data from the probe’s
XYZ coordinate system to an Earth (East-North-Up or ENU) coordinate system. The sensor used
is the Precision Navigation TCM2 Electric Compass Module.
•
Communication with the ADV is through an RS-232 interface. Although direct communica-
tion with the compass is not required under normal conditions, this interface lets you “talk”
directly with the compass (see §4.7.6).
•
When properly calibrated, the TCM2 provides accurate heading data (±2.0
°
) and accurate
pitch/roll data (±0.1
°
up to 50°).
•
There are limitations when using the ADV with a compass. Foremost is the 50° limit on
pitch/roll. Any pitch/roll beyond this will produce unpredictable results.
•
It is essential that the compass be correctly mounted relative to probe orientation to ensure
accurate compass data and the correct rotation of XYZ velocity data (see §4.7.2).
•
Care should be taken to limit how fast the orientation of the ADV changes because high-
frequency motion can distort compass data. The compass/tilt sensor has a maximum fre-
quency response of about 2 Hz.
•
Remember that the ADV measures velocity relative to its own speed – if the ADV is mov-
ing, the recorded velocity is the vector sum of the ADV velocity and the true water motion.
4.7.1. Testing Compass Operation
This section describes a simple procedure to verify basic compass operation performed using a
continuous display of compass data (heading, pitch, and roll). The display can be obtained using
the direct command interface, the DOS-based compass calibration software (
COMPCAL
), or the
Windows-based compass calibration software (
Compass
, part of
SonUtils
). If using the direct
command interface, the command
Compass CONT
gives a continuous output of heading, pitch,
and roll.
COMPCAL
and
Compass
(see
ADVField Software Manual
) provide continuously updated
graphical and tabular displays of compass data. Guidelines to running this test:
•
Be sure the compass is oriented vertically (§4.7.2) and the hardware orientation switches
(§4.2) match the compass installation.
•
The area around the ADV should be predominantly free of magnetic material.
•
Start the display of compass data.
•
Point the probe’s X-axis (if compass is mounted in conditioning module) or the underwater
canister’s X-axis (if the compass is mounted with the processing module) towards each
compass direction (north, south, east, and west).
•
Compare the compass output with the expected reading.
•
Test the pitch/roll sensors by tilting the ADV about the X-axis (roll) and Y-axis (pitch).
•
The compass and tilt data should appear reasonable to within the accuracy of the sensor
(
±
10
°
heading,
±
5
°
tilt). Large errors or a lack of response indicates a problem.
•
This is not a precise test of compass accuracy, but a means to evaluate basic performance.
•
For best results, we recommend performing a compass calibration before every deployment.
Содержание SonTek ADVField
Страница 2: ......
Страница 8: ...SonTek YSI iv ADVField Hydra Operation Manual September 1 2001 NOTES ...
Страница 12: ......
Страница 32: ...SonTek YSI 20 ADVField Operation Manual September 1 2001 ...
Страница 102: ...SonTek YSI 90 ADVField Operation Manual September 1 2001 ...
Страница 104: ...SonTek YSI 92 ADVField Operation Manual September 1 2001 ...
Страница 118: ...SonTek YSI 106 ADVField Operation Manual September 1 2001 ...
Страница 122: ...SonTek YSI 110 ADVField Operation Manual September 1 2001 ...
Страница 126: ...SonTek YSI 2 ADVField Operation Manual September 1 2001 ...
Страница 132: ...SonTek YSI vi ADVField Software Manual September 1 2001 NOTES ...
Страница 136: ...SonTek YSI ADVField Software Manual September 1 2001 4 ...
Страница 152: ...SonTek YSI ADVField Software Manual September 1 2001 20 ...
Страница 160: ...SonTek YSI ADVField Software Manual September 1 2001 28 ...
Страница 162: ...SonTek YSI ADVField Software Manual September 1 2001 30 ...
Страница 164: ...SonTek YSI ADVField Software Manual September 1 2001 32 ...
Страница 168: ...SonTek YSI ADVField Software Manual September 1 2001 36 ...
Страница 174: ...SonTek YSI ADVField Software Manual September 1 2001 42 ...
Страница 182: ...SonTek YSI ADVField Software Manual September 1 2001 50 ...
Страница 184: ...SonTek YSI ADVField Software Manual September 1 2001 52 Figure 15 ViewHydra Screen Shot ...
Страница 196: ...SonTek YSI ADVField Software Manual September 1 2001 64 ...
Страница 198: ...SonTek YSI ADVField Software Manual September 1 2001 66 ...
Страница 202: ......
Страница 206: ......
Страница 220: ...SonTek YSI Acoustic Doppler Velocimeter Principles of Operation September 1 2001 14 ...