SonTek/YSI
ADVField/Hydra Operation Manual (September 1, 2001)
55
•
For older, real-time ADVField systems (without the internal recorder or battery power):
o
The connector is a female 16-pin underwater mateable connector.
o
The bulkhead connector used is an Impulse MBH-16-FS.
o
The mating connector on the cable is an Impulse MIL-16-MP (with locking sleeve).
o
Table 19 gives the pin-out for this connector.
o
Note: This cable has been discontinued.
Table 19. Impulse MIL-16-MP Connector Pin-outs
Pin
# Signal Pin
# Signal Pin
# Signal
1 Vpower 7 Vz 13 Not
Used
2 Vx 8 Amp
14 Rx-
3 Vy 9
Ground
15
SyncOut
4
Ground
10
Data Out / Tx+
16
Ground
5
SyncIn
11
Data In / Tx-
6 Rx+ 12
Ground
The standard termination for the power/communication cable uses a DB9 connector for RS232
serial communication and a coaxial connector for power. The DB9 connector has the same pin-
out as the DB9 connector on the ADVField splash-proof housing (§4.3.1). The coaxial connector
uses the center conductor as positive and the outer conductor as negative. Contact SonTek for
more information about custom cable wiring and access to the auxiliary input/output signals.
4.3.3. External Sensor Connector
Systems with external sensors may use different cables/connectors depending on the sensor being
used. Two of the more common connector pin-outs are shown in the following tables. If you are
unsure of your connector type, contact SonTek.
Table 20. Sensor Connector Pin-outs (8-pin, underwater, male)
Pin #
Signal
Pin #
Signal
Pin #
Signal
1
Sensor power out
4
Ext Sensor Chan 1
7
Sync In
2
Ext Serial Data In
5
Ext Sensor Chan 2
8
Ground
3
Ext Serial Data Out
6
Sync Out
Table 21. Sensor Connector Pin-outs (16-pin, underwater, female)
Pin #
Signal
Pin #
Signal
Pin #
Signal
1
Not used
7
Paros temperature
13
Not used
2
Ext Sensor Chan 1
8
Paros pressure
14
Not used
3
Ext Sensor Chan 2
9
Ground
15
Ext Serial Data Out
4
Sync Out
10
Not used
16
Ground
5
Sync In
11
Not used
6
Not used
12
Ext Serial Data In
Содержание SonTek ADVField
Страница 2: ......
Страница 8: ...SonTek YSI iv ADVField Hydra Operation Manual September 1 2001 NOTES ...
Страница 12: ......
Страница 32: ...SonTek YSI 20 ADVField Operation Manual September 1 2001 ...
Страница 102: ...SonTek YSI 90 ADVField Operation Manual September 1 2001 ...
Страница 104: ...SonTek YSI 92 ADVField Operation Manual September 1 2001 ...
Страница 118: ...SonTek YSI 106 ADVField Operation Manual September 1 2001 ...
Страница 122: ...SonTek YSI 110 ADVField Operation Manual September 1 2001 ...
Страница 126: ...SonTek YSI 2 ADVField Operation Manual September 1 2001 ...
Страница 132: ...SonTek YSI vi ADVField Software Manual September 1 2001 NOTES ...
Страница 136: ...SonTek YSI ADVField Software Manual September 1 2001 4 ...
Страница 152: ...SonTek YSI ADVField Software Manual September 1 2001 20 ...
Страница 160: ...SonTek YSI ADVField Software Manual September 1 2001 28 ...
Страница 162: ...SonTek YSI ADVField Software Manual September 1 2001 30 ...
Страница 164: ...SonTek YSI ADVField Software Manual September 1 2001 32 ...
Страница 168: ...SonTek YSI ADVField Software Manual September 1 2001 36 ...
Страница 174: ...SonTek YSI ADVField Software Manual September 1 2001 42 ...
Страница 182: ...SonTek YSI ADVField Software Manual September 1 2001 50 ...
Страница 184: ...SonTek YSI ADVField Software Manual September 1 2001 52 Figure 15 ViewHydra Screen Shot ...
Страница 196: ...SonTek YSI ADVField Software Manual September 1 2001 64 ...
Страница 198: ...SonTek YSI ADVField Software Manual September 1 2001 66 ...
Страница 202: ......
Страница 206: ......
Страница 220: ...SonTek YSI Acoustic Doppler Velocimeter Principles of Operation September 1 2001 14 ...