Case F: Damaged receiver arm (Figure 9)
If the sampling volume in one receiver is offset along the horizontal axis, the receiver arm has
been physically bent. Often, bent probes can be repaired quickly and inexpensively by unbending
the arm and recalibrating the probe. Contact SonTek if you have a damaged probe for details
about repair options.
Case G: Excessive noise
If the noise is more than 10 counts above its normal level, it may be a sign that something is
wrong with the probe, the signal-conditioning module, or the cable. The noise level may also be
higher than normal because of external (electromagnetic) interference, or because water has con-
densed on the receiver electronics. Coil any extra high frequency probe cable to minimize the
noise and keep the electronics dry by putting a fresh bag of desiccant inside the conditioning
module.
Case H: High signal strength past the boundary (Figure 10)
If the signal strength does not fall off rapidly past the boundary, the additional acoustic noise can
potentially increase the noise in velocity measurements. This additional acoustic noise is most
commonly seen in small tanks, particularly when built from acrylic or glass. To improve the
situation, cover the boundary with a piece of cloth, a rubber mat, or a sheet of acoustically ab-
sorbing material.
Содержание SonTek ADVField
Страница 2: ......
Страница 8: ...SonTek YSI iv ADVField Hydra Operation Manual September 1 2001 NOTES ...
Страница 12: ......
Страница 32: ...SonTek YSI 20 ADVField Operation Manual September 1 2001 ...
Страница 102: ...SonTek YSI 90 ADVField Operation Manual September 1 2001 ...
Страница 104: ...SonTek YSI 92 ADVField Operation Manual September 1 2001 ...
Страница 118: ...SonTek YSI 106 ADVField Operation Manual September 1 2001 ...
Страница 122: ...SonTek YSI 110 ADVField Operation Manual September 1 2001 ...
Страница 126: ...SonTek YSI 2 ADVField Operation Manual September 1 2001 ...
Страница 132: ...SonTek YSI vi ADVField Software Manual September 1 2001 NOTES ...
Страница 136: ...SonTek YSI ADVField Software Manual September 1 2001 4 ...
Страница 152: ...SonTek YSI ADVField Software Manual September 1 2001 20 ...
Страница 160: ...SonTek YSI ADVField Software Manual September 1 2001 28 ...
Страница 162: ...SonTek YSI ADVField Software Manual September 1 2001 30 ...
Страница 164: ...SonTek YSI ADVField Software Manual September 1 2001 32 ...
Страница 168: ...SonTek YSI ADVField Software Manual September 1 2001 36 ...
Страница 174: ...SonTek YSI ADVField Software Manual September 1 2001 42 ...
Страница 182: ...SonTek YSI ADVField Software Manual September 1 2001 50 ...
Страница 184: ...SonTek YSI ADVField Software Manual September 1 2001 52 Figure 15 ViewHydra Screen Shot ...
Страница 196: ...SonTek YSI ADVField Software Manual September 1 2001 64 ...
Страница 198: ...SonTek YSI ADVField Software Manual September 1 2001 66 ...
Страница 202: ......
Страница 206: ......
Страница 220: ...SonTek YSI Acoustic Doppler Velocimeter Principles of Operation September 1 2001 14 ...