![YSI SonTek ADVField Скачать руководство пользователя страница 49](http://html.mh-extra.com/html/ysi/sontek-advfield/sontek-advfield_technical-documentation-manual_915334049.webp)
SonTek/YSI
ADVField/Hydra Operation Manual (September 1, 2001)
37
SamplesPerBurst or SPB [d] [d] [d]
•
Default parameters:
1200 0 0
•
Parameter range:
1
to
32000
•
Without parameter: returns its current settings for each burst type.
•
When used with from one to three valid parameters, sets the number of samples per burst for
each burst type. See §1.3 for a description of burst sampling.
•
For autonomous deployments, the ADVField supports up to three samples per burst values
for three burst types. Samples per burst are not used during real-time data collection.
•
To select multiple-burst sampling, enter a separate samples-per-burst for each burst type de-
sired. Example:
SPB 24 600 7500
sets the samples per burst for burst type #1 to 24, for
burst type #2 to 600, for burst type #3 to 7500. If only one or two input arguments are speci-
fied, the values for the other burst types are set to 0, disabling that burst type.
RecordAmpCorr [YES|NO] [YES|NO] [YES|NO]
•
Default parameters:
YES YES YES
•
Without parameter: returns its current settings for each burst type.
•
When used with from one to three valid parameters, tells the ADV whether to record ampli-
tude and correlation data. See §1.3 for a description of burst sampling.
○
YES
tells the ADV to record amplitude and correlation data for each receiver with each
sample for the selected burst types.
YES
increases the size of each sample by 4 bytes.
○
NO
tells the ADV to record only the mean amplitude (from all 3 receivers) and the mean
correlation (from all 3 receivers) with each sample for the selected burst types.
•
This parameter is used only in deployment mode; it has no affect on real-time data collection.
•
If only one or two input arguments are specified, it does not change the settings for the non-
specified burst types.
•
Appendix 2 has a description of the data recorded based on the
Record
command settings.
RecordCompass [YES|NO] [YES|NO] [YES|NO]
•
Default parameters:
○
No compass/tilt sensor installed:
NO NO NO
○
With compass/tile sensor installed:
YES YES YES
•
Without parameter: returns its current settings for each burst type.
•
When used with from one to three valid parameters, tells the ADV whether to record com-
pass/tilt sensor data. See §1.3 for a description of burst sampling.
○
YES
tells the ADV to record compass/tilt sensor data (heading, pitch, roll) for each veloc-
ity sample for the selected burst types.
YES
increases the size of each sample by 6 bytes.
○
NO
tells the ADV not to record compass/tilt sensor data for the selected burst types.
○
Regardless of this setting, the ADV will always record velocity data in either XYZ or
ENU coordinates based on the setting of the
CoordSystem
setup parameter (see §3.6).
•
This parameter is used only in deployment mode; it has no affect on real-time data collection.
•
If only one or two input arguments are specified, it does not change the settings for the non-
specified burst types.
•
Appendix 2 has a description of the data recorded based on the
Record
command settings.
Содержание SonTek ADVField
Страница 2: ......
Страница 8: ...SonTek YSI iv ADVField Hydra Operation Manual September 1 2001 NOTES ...
Страница 12: ......
Страница 32: ...SonTek YSI 20 ADVField Operation Manual September 1 2001 ...
Страница 102: ...SonTek YSI 90 ADVField Operation Manual September 1 2001 ...
Страница 104: ...SonTek YSI 92 ADVField Operation Manual September 1 2001 ...
Страница 118: ...SonTek YSI 106 ADVField Operation Manual September 1 2001 ...
Страница 122: ...SonTek YSI 110 ADVField Operation Manual September 1 2001 ...
Страница 126: ...SonTek YSI 2 ADVField Operation Manual September 1 2001 ...
Страница 132: ...SonTek YSI vi ADVField Software Manual September 1 2001 NOTES ...
Страница 136: ...SonTek YSI ADVField Software Manual September 1 2001 4 ...
Страница 152: ...SonTek YSI ADVField Software Manual September 1 2001 20 ...
Страница 160: ...SonTek YSI ADVField Software Manual September 1 2001 28 ...
Страница 162: ...SonTek YSI ADVField Software Manual September 1 2001 30 ...
Страница 164: ...SonTek YSI ADVField Software Manual September 1 2001 32 ...
Страница 168: ...SonTek YSI ADVField Software Manual September 1 2001 36 ...
Страница 174: ...SonTek YSI ADVField Software Manual September 1 2001 42 ...
Страница 182: ...SonTek YSI ADVField Software Manual September 1 2001 50 ...
Страница 184: ...SonTek YSI ADVField Software Manual September 1 2001 52 Figure 15 ViewHydra Screen Shot ...
Страница 196: ...SonTek YSI ADVField Software Manual September 1 2001 64 ...
Страница 198: ...SonTek YSI ADVField Software Manual September 1 2001 66 ...
Страница 202: ......
Страница 206: ......
Страница 220: ...SonTek YSI Acoustic Doppler Velocimeter Principles of Operation September 1 2001 14 ...