SonTek/YSI
12
ADVField Operation Manual (September 1, 2001)
ADV enters a low power standby mode when not collecting data). Using the above parameters,
the ADV generates between 230 and 630 KB of data per day.
Users are often interested in different aspects of the velocity field that require different sampling
schemes. Examples might include mean currents (requiring slow sampling rate), waves (requir-
ing a moderate sampling rate; 2-5 Hz), and turbulence (requiring a rapid sampling rate;
≈
25 Hz).
The autonomous ADV has been designed to allow up to three different burst types to be collected
in a single deployment. Each burst type has independent values for burst interval, sampling rate,
and samples per burst. To avoid conflicts, these parameters are subject to the following rules.
1. The burst intervals for types 2 and 3 must be integer multiples of the burst interval for type 1.
2. The sum of the three sampling times must be at least 15 seconds less than the burst interval for
type 1 (to allow an overhead of 5 seconds for each burst).
The last type of sampling in Figure 7 shows multiple-burst sampling; lines of different heights
represent different burst types. Multiple burst sampling is best described with an example.
Suppose you are interested in mean currents, waves, and turbulence for a particular deployment.
For mean currents, you might sample at 0.2 Hz (every 5 seconds) for a period of 2 minutes, and
repeat the sampling every 20 minutes. For waves, you might sample at 4 Hz for a period of 5
minutes, and repeat the sampling every hour. For turbulence, you might sample at 25 Hz for a pe-
riod of 5 minutes, and repeat the sampling every two hours. To do this, you could set the parame-
ters shown in Table 2.
Note: Changing the sampling rate does not effect power consumption. As described earlier, the
ADV pings as rapidly as it can, given a certain velocity range and configuration settings,
regardless of the sampling rate. However, changing the velocity range does change the
power consumption.
To calculate the total duty cycle, add the duty cycle from each burst (for this example, it is about
23%). When calculating memory requirements, the burst interval as well as the amount of data
must to be taken into account. The example shown here would generate 0.8 to 2.3 MB of data per
day, depending on what parameters are recorded (see §3.7). All data is stored in one file, and the
data conversion software accesses each burst type independently to simplify its presentation and
to separate the burst types for analysis (see
ADVField Software Manual
for details). Multiple-
burst sampling supports a wide variety of sampling strategies in a reliable and easy to use format.
Burst Type 1:
Mean Currents
Burst Type 2:
Waves
Burst Type 3:
Turbulence
Sampling Rate (Hz)
0.2
2.0
25.0
Samples per Burst
24
600
7500
Burst Interval (s)
1200
3600
7200
Sample Time (s)
120
300
300
Burst Duty Cycle (%)
10.4%
8.5%
4.3%
Kbytes per Burst
0.2 - 0.5
4.8 - 13.2
60 - 160
Table 2. Example Parameters for Multiple-Burst Sampling Strategy
Содержание SonTek ADVField
Страница 2: ......
Страница 8: ...SonTek YSI iv ADVField Hydra Operation Manual September 1 2001 NOTES ...
Страница 12: ......
Страница 32: ...SonTek YSI 20 ADVField Operation Manual September 1 2001 ...
Страница 102: ...SonTek YSI 90 ADVField Operation Manual September 1 2001 ...
Страница 104: ...SonTek YSI 92 ADVField Operation Manual September 1 2001 ...
Страница 118: ...SonTek YSI 106 ADVField Operation Manual September 1 2001 ...
Страница 122: ...SonTek YSI 110 ADVField Operation Manual September 1 2001 ...
Страница 126: ...SonTek YSI 2 ADVField Operation Manual September 1 2001 ...
Страница 132: ...SonTek YSI vi ADVField Software Manual September 1 2001 NOTES ...
Страница 136: ...SonTek YSI ADVField Software Manual September 1 2001 4 ...
Страница 152: ...SonTek YSI ADVField Software Manual September 1 2001 20 ...
Страница 160: ...SonTek YSI ADVField Software Manual September 1 2001 28 ...
Страница 162: ...SonTek YSI ADVField Software Manual September 1 2001 30 ...
Страница 164: ...SonTek YSI ADVField Software Manual September 1 2001 32 ...
Страница 168: ...SonTek YSI ADVField Software Manual September 1 2001 36 ...
Страница 174: ...SonTek YSI ADVField Software Manual September 1 2001 42 ...
Страница 182: ...SonTek YSI ADVField Software Manual September 1 2001 50 ...
Страница 184: ...SonTek YSI ADVField Software Manual September 1 2001 52 Figure 15 ViewHydra Screen Shot ...
Страница 196: ...SonTek YSI ADVField Software Manual September 1 2001 64 ...
Страница 198: ...SonTek YSI ADVField Software Manual September 1 2001 66 ...
Страница 202: ......
Страница 206: ......
Страница 220: ...SonTek YSI Acoustic Doppler Velocimeter Principles of Operation September 1 2001 14 ...