SonTek/YSI
ADVField/Hydra Operation Manual (September 1, 2001)
17
1. Determine the number of bytes recorded per sample (§3.7).
2. Calculate the number of bytes to be recorded per burst (below).
3. Calculate the total file size based on expected number of bursts (below).
Number of bytes per burst
is calculated based on the bytes per sample (SAMP_SIZE) from
Step 1 above, the number of samples per burst (N_SAMP), and whether burst statistics are re-
corded. If burst statistics are not recorded, use the formula:
BURST_SIZE = 62 + (N_SAMP*SAMP_SIZE)
If burst statistics are recorded, use the formula:
BURST_SIZE = 98 + (N_SAMP*SAMP_SIZE)
Total file size
is the number of bursts multiplied by the bytes for each burst plus the file header
(441 bytes). When using multiple-burst sampling, you must calculate the burst size and number
of bursts for each burst type. See §A2.2 for a detailed description of the data file format.
The ADV recorder is available in sizes ranging from 20 to 880 MB, and therefore can accommo-
date a wide range of sampling strategies and deployment lengths. The ADVField recorder is di-
vided into blocks of 64 KB each; only one data file can be written in one block, although one
data file can occupy multiple blocks.
Bear in mind that the recorder also has a maximum number of
files
that can be stored. This num-
ber is computed by dividing the recorder size by 64 KB. If you record a large number of small
data files (less than 64 KB per file), the recorder will fill before reaching the rated maximum ca-
pacity. Be sure to format the recorder to erase all files before any extended deployment.
2.4.3. Preparing the ADVField for Autonomous Operation
This section assumes that the ADVField system includes the internal recorder in the same un-
derwater housing as the ADVField processing module. The instructions below assume that you
will use an external power supply to set up your system before deployment (or testing). If your
system is equipped with the internal battery option, the deployment (or testing) setup can be done
from battery power simply by not connecting the external power supply.
1. Determine the power requirements (§2.4.1) and memory requirements (§2.4.2) for your ap-
plication. Also, review the operational considerations presented in Section 5.
2. Connect the high-frequency probe cable to the underwater canister. Connect the other end of
the high-frequency cable to the probe. Connect the power and communication cable to the
underwater canister. See §4.3 for cables and connector details.
3. Connect the DB9 connector on the power and communication cable to an available serial port
on your computer. Note: For RS422/485 systems, use the RS232/422 adaptor provided with
the system.
4. Connect the coaxial power connector on the power and communication cable to the power
supply included with the system. Plug the supply into an available 110/220 VAC outlet. The
external power supply is used if its voltage is greater than the internal battery packs. The
standard ADV battery packs are wired for 18 VDC; the power supply included with the sys-
tem supplies 24 VDC. Note: If desired, you can skip this step if you want to set up the ADV
for deployment (or testing) using the battery packs instead of an external power supply.
5. If you are using the system directly out of the box, go to Step 6; otherwise, you may need to
edit the ASCII text file
ADVPROBE.DEF
. This file should contain one line that looks similar to
PROBE 0 A001.PRO
. The number
0
indicates the address of the ADV to which you are estab-
Содержание SonTek ADVField
Страница 2: ......
Страница 8: ...SonTek YSI iv ADVField Hydra Operation Manual September 1 2001 NOTES ...
Страница 12: ......
Страница 32: ...SonTek YSI 20 ADVField Operation Manual September 1 2001 ...
Страница 102: ...SonTek YSI 90 ADVField Operation Manual September 1 2001 ...
Страница 104: ...SonTek YSI 92 ADVField Operation Manual September 1 2001 ...
Страница 118: ...SonTek YSI 106 ADVField Operation Manual September 1 2001 ...
Страница 122: ...SonTek YSI 110 ADVField Operation Manual September 1 2001 ...
Страница 126: ...SonTek YSI 2 ADVField Operation Manual September 1 2001 ...
Страница 132: ...SonTek YSI vi ADVField Software Manual September 1 2001 NOTES ...
Страница 136: ...SonTek YSI ADVField Software Manual September 1 2001 4 ...
Страница 152: ...SonTek YSI ADVField Software Manual September 1 2001 20 ...
Страница 160: ...SonTek YSI ADVField Software Manual September 1 2001 28 ...
Страница 162: ...SonTek YSI ADVField Software Manual September 1 2001 30 ...
Страница 164: ...SonTek YSI ADVField Software Manual September 1 2001 32 ...
Страница 168: ...SonTek YSI ADVField Software Manual September 1 2001 36 ...
Страница 174: ...SonTek YSI ADVField Software Manual September 1 2001 42 ...
Страница 182: ...SonTek YSI ADVField Software Manual September 1 2001 50 ...
Страница 184: ...SonTek YSI ADVField Software Manual September 1 2001 52 Figure 15 ViewHydra Screen Shot ...
Страница 196: ...SonTek YSI ADVField Software Manual September 1 2001 64 ...
Страница 198: ...SonTek YSI ADVField Software Manual September 1 2001 66 ...
Страница 202: ......
Страница 206: ......
Страница 220: ...SonTek YSI Acoustic Doppler Velocimeter Principles of Operation September 1 2001 14 ...