SonTek/YSI
ADVField/Hydra Operation Manual (September 1, 2001)
43
○
If the frequency reading is reasonable, but the corresponding pressure is not, this indi-
cates that the appropriate calibration is not loaded into the processor. Use the
LDDRUCK
program (see
ADVField Software Manual
) to load the
DRUCK
calibration.
Show DRUCK
or
Show PAROSFREQ
•
Displays
DRUCK
/
PAROSFREQ
calibration data in the following format:
Druck pressure sensor calibration data
--------------------------------------
SN:
1276950
F0:
1.200e+07
X:
8.403
Y:
5.602
T0: -1.001
D0:
6.202
DT: -2.200
K00:
1.503
K01: -2.6693000e-02
K10:
1.600
K11: -1.0395800e-05
K20:
3.9933779e-04
K21: -4.9470601e-09
K30:
8.0327140e-08
K31: -1.4093000e-12
K40:
1.7388157e-11
K41:
2.2691999e-16
K50:
2.5953699e-15
K51:
1.3877999e-19
K02:
3.1005999e-05
K03:
2.6474001e-07
K12:
1.8332999e-08
K13:
8.2459997e-11
K22:
1.6572001e-11
K23: -1.7079999e-14
K32: -7.7020003e-15
K33: -9.9870002e-17
K42: -4.0642002e-18
K43: -1.0678000e-19
K52:
5.3660002e-21
K53: -2.8249999e-23
3.9.5. External Sensor Commands – LISST-100 Sensor
These commands apply to systems that have an external LISST sensor installed (§4.12). The
LISST-100 is a Laser Interferometry and Transmissometry particle distribution measurement de-
vice designed by Sequoia Scientific.
LisstSensorInstalled Set [YES|NO]
•
Selects or disables the LISST sensor (§4.12).
•
Set at the factory. Does not need to be changed except in some special cases.
•
The LISST sensor is a serial instrument that uses the second COM port on the ADVField’s
CPU board. As such, other serial devices (e.g., Paros, MicroCat CTD) must be disabled.
•
When set to YES, LisstBufferSize is automatically set to 80 bytes (default).
LisstBufferSize
or
LBS [d]
•
Default parameter:
80
bytes
•
Without parameter: returns its current setting.
•
When used with a valid parameter, sets the size of the LISST sample buffer.
Содержание SonTek ADVField
Страница 2: ......
Страница 8: ...SonTek YSI iv ADVField Hydra Operation Manual September 1 2001 NOTES ...
Страница 12: ......
Страница 32: ...SonTek YSI 20 ADVField Operation Manual September 1 2001 ...
Страница 102: ...SonTek YSI 90 ADVField Operation Manual September 1 2001 ...
Страница 104: ...SonTek YSI 92 ADVField Operation Manual September 1 2001 ...
Страница 118: ...SonTek YSI 106 ADVField Operation Manual September 1 2001 ...
Страница 122: ...SonTek YSI 110 ADVField Operation Manual September 1 2001 ...
Страница 126: ...SonTek YSI 2 ADVField Operation Manual September 1 2001 ...
Страница 132: ...SonTek YSI vi ADVField Software Manual September 1 2001 NOTES ...
Страница 136: ...SonTek YSI ADVField Software Manual September 1 2001 4 ...
Страница 152: ...SonTek YSI ADVField Software Manual September 1 2001 20 ...
Страница 160: ...SonTek YSI ADVField Software Manual September 1 2001 28 ...
Страница 162: ...SonTek YSI ADVField Software Manual September 1 2001 30 ...
Страница 164: ...SonTek YSI ADVField Software Manual September 1 2001 32 ...
Страница 168: ...SonTek YSI ADVField Software Manual September 1 2001 36 ...
Страница 174: ...SonTek YSI ADVField Software Manual September 1 2001 42 ...
Страница 182: ...SonTek YSI ADVField Software Manual September 1 2001 50 ...
Страница 184: ...SonTek YSI ADVField Software Manual September 1 2001 52 Figure 15 ViewHydra Screen Shot ...
Страница 196: ...SonTek YSI ADVField Software Manual September 1 2001 64 ...
Страница 198: ...SonTek YSI ADVField Software Manual September 1 2001 66 ...
Страница 202: ......
Страница 206: ......
Страница 220: ...SonTek YSI Acoustic Doppler Velocimeter Principles of Operation September 1 2001 14 ...