![YSI SonTek ADVField Скачать руководство пользователя страница 223](http://html.mh-extra.com/html/ysi/sontek-advfield/sontek-advfield_technical-documentation-manual_915334223.webp)
SonTek/YSI
Using Frequency Pressure Sensors with SonTek/YSI Systems (December 2001)
3
3.2.
The Paroscientific Digiquartz (PAROSFREQ)
User requirements for deeper ratings (compared to
DRUCK
) and faster response (compared to
PAROS
) convinced SonTek to offer the Paroscientific digiquartz sensor with frequency output
(
PAROSFREQ
). Functionally, the interface for the
PAROSFREQ
is identical to the interface for the
DRUCK
, as both are frequency-based sensors. The
PAROSFREQ
sensors are available for ADP,
PC-ADP, and Hydra systems. The differences, when compared with a strain-gage sensor, are as
follows:
•
The
PAROSFREQ
is an absolute sensor. It reads 1 dbar in air; a strain gage reports 0 dbars.
•
Requires a special interface to sample the frequency signal.
•
Requires a special frequency splitter cable.
•
Sampling rate is limited to 10 Hz (ADVs and Hydras) because of the time required to sam-
ple the frequency signal with sufficient precision.
•
Requires a calibration (specific to each sensor) to be loaded into the processor (system CPU
or a user PC) to obtain valid pressure readings.
•
Available in 50-psia (
≈
22 dbar strain gage) and 100-psia (
≈
60-m strain gage) pressure ratings.
The structure of the calibration file corresponding to the
PAROSFREQ
sensor is similar to the
DRUCK
. For historic reasons,
PAROSFREQ
calibration file uses the .drk extension.
•
Pressure Sensor Serial Number
•
Oscillator Frequency (in Hz)
•
U0
•
N/A
•
N/A
•
N/A
•
N/A
•
Y1 Y2 Y3 N/A
•
C1 C2 C3 N/A
•
D1 D2 N/A N/A
•
T1 T2 T3 T4
•
N/A N/A N/A N/A
•
N/A N/A N/A N/A
Example – Paroscientific calibration file
p68888.drk
:
P68888
12.006
5.880670
0.0
0.0
0.0
0.0
-3898.415
-10940.31
0.0
0.0
224.5927
6.013600
-245.2861
0.0
0.041219
0.0
0.0
0.0
27.94067
0.859765
20.65914
26.99304
0.0
0.0
0.0
0.0
0.0
0.0
0.0
0.0
Note: The U, Y, C, D, and T coefficients can
be found on the Paroscientific calibration sheet.
Содержание SonTek ADVField
Страница 2: ......
Страница 8: ...SonTek YSI iv ADVField Hydra Operation Manual September 1 2001 NOTES ...
Страница 12: ......
Страница 32: ...SonTek YSI 20 ADVField Operation Manual September 1 2001 ...
Страница 102: ...SonTek YSI 90 ADVField Operation Manual September 1 2001 ...
Страница 104: ...SonTek YSI 92 ADVField Operation Manual September 1 2001 ...
Страница 118: ...SonTek YSI 106 ADVField Operation Manual September 1 2001 ...
Страница 122: ...SonTek YSI 110 ADVField Operation Manual September 1 2001 ...
Страница 126: ...SonTek YSI 2 ADVField Operation Manual September 1 2001 ...
Страница 132: ...SonTek YSI vi ADVField Software Manual September 1 2001 NOTES ...
Страница 136: ...SonTek YSI ADVField Software Manual September 1 2001 4 ...
Страница 152: ...SonTek YSI ADVField Software Manual September 1 2001 20 ...
Страница 160: ...SonTek YSI ADVField Software Manual September 1 2001 28 ...
Страница 162: ...SonTek YSI ADVField Software Manual September 1 2001 30 ...
Страница 164: ...SonTek YSI ADVField Software Manual September 1 2001 32 ...
Страница 168: ...SonTek YSI ADVField Software Manual September 1 2001 36 ...
Страница 174: ...SonTek YSI ADVField Software Manual September 1 2001 42 ...
Страница 182: ...SonTek YSI ADVField Software Manual September 1 2001 50 ...
Страница 184: ...SonTek YSI ADVField Software Manual September 1 2001 52 Figure 15 ViewHydra Screen Shot ...
Страница 196: ...SonTek YSI ADVField Software Manual September 1 2001 64 ...
Страница 198: ...SonTek YSI ADVField Software Manual September 1 2001 66 ...
Страница 202: ......
Страница 206: ......
Страница 220: ...SonTek YSI Acoustic Doppler Velocimeter Principles of Operation September 1 2001 14 ...