![YSI SonTek ADVField Скачать руководство пользователя страница 150](http://html.mh-extra.com/html/ysi/sontek-advfield/sontek-advfield_technical-documentation-manual_915334150.webp)
SonTek/YSI
ADVField Software Manual (September 1, 2001)
18
File names must be given without an extension. The programs assume the ADV binary data file
has the extension
.ADV
and assigns the output file an extension based on the type of data being
extracted (extensions are
.AMP
,
.COR
,
.HPR
,
.SEN
,
.SNR
, and
.VEL
). If no output file is specified,
the programs assign the same root name as the binary data file.
Note: The
[probe #]
parameter is only used with ADVLab systems, which are not discussed in
this document.
The
[first sample]
and
[last sample]
permit you to extract a limited portion of the data. If
they are not specified, the programs extract all samples contained in the binary data file.
For example,
GETVEL TEST1 TEST 100 199
extracts the velocity data for samples 100 through 199 in the binary data file
TEST1.ADV
and
sends the output to ASCII file
TEST.VEL
.
Similarly, the command
GETSNR TEST1 TEST
extracts the SNR data from all samples from binary data file
TEST1.ADV
and sends the output to
ASCII file
TEST.SNR
.
These programs generate output files in a tabular ASCII format with one line per sample. With
the exception of the programs
GETHPR
and
GETSEN
(which are discussed at the end of this sec-
tion), the output files use the following format for each line.
T Event Probe# DATA1 DATA2 DATA3 Probe# DATA1 DATA2 DATA3
where
T
is the time in seconds from the start of the first sample, and
Event
is the value of the
event counter (for marking points during the data file using <F7>) at the time the sample was col-
lected. The rest of the line contains the probe number followed by its corresponding data. Table 2
summarizes the values and units output by the data sample conversion programs.
Program
DATA1 DATA2 DATA3 UNITS
GETAMP
Amplitude
Beam 1
Amplitude
Beam 2
Amplitude
Beam 3
Counts
(1 count = 0.43 dB)
GETCOR
Correlation
Beam 1
Correlation
Beam 2
Correlation
Beam 3
Percentage
(0-100%)
GETSNR
SNR Beam 1
SNR Beam 2
SNR Beam 3
dB
GETVEL
Vx or Veast
Vy or Vnorth
Vz or Vup
cm/s or ft/s
The programs
GETHPR
and
GETSEN
extract data from the optional sensors on the ADVField. The
first three values on each line are time (in seconds from the start of the file), event counter, and
probe number. If the compass/tilt sensor is installed, both
GETHPR
and
GETSEN
extract heading,
pitch, and roll (in degrees) as the next three data values. If either the temperature or pressure sen-
sor is installed, but the compass/tilt sensor is not,
GETSEN
extracts temperature (in either
°
C
or
°
F) and pressure (in either dBar or psi) as the next two data values. If compass/tilt, tempera-
ture, and pressure are installed,
GETSEN
extracts five data values in the following order: heading,
pitch, roll, temperature, and pressure.
Table 2 . Conversion Programs Data Output Summary
Содержание SonTek ADVField
Страница 2: ......
Страница 8: ...SonTek YSI iv ADVField Hydra Operation Manual September 1 2001 NOTES ...
Страница 12: ......
Страница 32: ...SonTek YSI 20 ADVField Operation Manual September 1 2001 ...
Страница 102: ...SonTek YSI 90 ADVField Operation Manual September 1 2001 ...
Страница 104: ...SonTek YSI 92 ADVField Operation Manual September 1 2001 ...
Страница 118: ...SonTek YSI 106 ADVField Operation Manual September 1 2001 ...
Страница 122: ...SonTek YSI 110 ADVField Operation Manual September 1 2001 ...
Страница 126: ...SonTek YSI 2 ADVField Operation Manual September 1 2001 ...
Страница 132: ...SonTek YSI vi ADVField Software Manual September 1 2001 NOTES ...
Страница 136: ...SonTek YSI ADVField Software Manual September 1 2001 4 ...
Страница 152: ...SonTek YSI ADVField Software Manual September 1 2001 20 ...
Страница 160: ...SonTek YSI ADVField Software Manual September 1 2001 28 ...
Страница 162: ...SonTek YSI ADVField Software Manual September 1 2001 30 ...
Страница 164: ...SonTek YSI ADVField Software Manual September 1 2001 32 ...
Страница 168: ...SonTek YSI ADVField Software Manual September 1 2001 36 ...
Страница 174: ...SonTek YSI ADVField Software Manual September 1 2001 42 ...
Страница 182: ...SonTek YSI ADVField Software Manual September 1 2001 50 ...
Страница 184: ...SonTek YSI ADVField Software Manual September 1 2001 52 Figure 15 ViewHydra Screen Shot ...
Страница 196: ...SonTek YSI ADVField Software Manual September 1 2001 64 ...
Страница 198: ...SonTek YSI ADVField Software Manual September 1 2001 66 ...
Страница 202: ......
Страница 206: ......
Страница 220: ...SonTek YSI Acoustic Doppler Velocimeter Principles of Operation September 1 2001 14 ...