This “Operation Manual” describes information on 

how to operate “PANOURA 18S”, a digital panorama 

tomographic X-ray diagnostic apparatus for dental 

use, and precaustions for use.

•  To avoid any unexpected accident, failure, or 

damage to the system, read this “Operation 

Manual” prior to system installation.

•  Be sure to keep this manual after having read 

through it.

PANOURA 18S

Operation Manual

Содержание PANOURA 18S

Страница 1: ...ama tomographic X ray diagnostic apparatus for dental use and precaustions for use To avoid any unexpected accident failure or damage to the system read this Operation Manual prior to system installat...

Страница 2: ......

Страница 3: ...tion or opinion about the manual feel free to contact your dealer Should any nonconsecutive numbering or page missing is found the defective manual will be replaced with a new one Reproduction or repr...

Страница 4: ...ded Use 19 1 2 Main label sealing 19 1 3 Names and functions of units and their parts 23 1 3 1 Main body 23 1 3 2 Names and functions of control panel switches and indicators 1 25 1 3 3 Names and func...

Страница 5: ...r 67 2 3 10 Shutting off the power supply 68 2 4 Procedures of Cephalometric Radiography 69 2 4 1 Turning the power switch on 69 2 4 2 Activating Image Creator from the Viewer 69 2 4 3 Setting the Ima...

Страница 6: ...quired Data RAW Data 99 4 7 Tomotgraphic Adjustment 99 4 8 Display Adjustment 101 4 9 Image Export 102 5 User Program Mode 103 6 Recovering from failures 107 6 1 Problems and their corrective measures...

Страница 7: ...Standing position base mount short type with an optional wide base Panorama type 133 11 3 9 Standing position base mount long type with an optional wide base Panorama type 134 11 3 10 Standing positio...

Страница 8: ...eration Manual 11 6 Spatial resolution MTF 146 11 7 Anode cooling curve and heat content of X ray tube assembly 147 12 Qualification Assurance QA Procedure 149 12 1 Structure of the phantom 149 12 2 Q...

Страница 9: ...vere injury or death serious damage to the equipment and its surrounding proprietary e g totally lost machine or fire WARNING This is used to indicate an indirect or potential danger that could cause...

Страница 10: ...r or radio controlled transmitter 7 Be sure to plug a medical ground cable into a ground outlet prior to use of PANOURA 18S Don t plug the cable into any other kind of outlet than that for exclusive u...

Страница 11: ...ccessory than those we designated 10 Keep the patient out from the equipment until the arm unit of the equipment has re turned to its start position Otherwise the patient may be caught by the arm unit...

Страница 12: ...he image quality 26 Set up the monitor in position so as to avoid any direct light reflection from any lighting inside or outside thus ensuring the best image quality and viewability 27 Do not operate...

Страница 13: ...ng as the equipment is properly maintained and inspected Based on Company data Repair functional part supply period Supply of the replacement performance parts for this product parts required to maint...

Страница 14: ...e manual NOTE Provides general information on operations independent of safety precautions Suppl Provides supplementary information on operations and functions independent of safety precautions See al...

Страница 15: ...ronment specified below The customer or the user of the Panoura18S should assure that is used in such an environment Immunity Test IEC 60601 Test Level Compliance Level Electromagnetic Environment gui...

Страница 16: ...agnetic site surveya should be less than the com pliance level in each frequency rangeb Interference may occur in the vicinity of equipment marked with the following symbol Radiated RF IEC 61000 4 3 3...

Страница 17: ...MHz d 1 2 P 80 MHz to 800 MHz d 1 2 P 800 MHz to 2 5 GHz d 2 3 P 0 01 0 12 0 12 0 23 0 1 0 38 0 38 0 73 1 1 2 1 2 2 3 10 3 8 3 8 7 3 100 12 12 23 For transmitter rated at a maximum output power not l...

Страница 18: ...to 75 no condensation 5 Areas subject to steam 6 Areas subject to water drops 7 Areas with much dust or sand 8 Areas with much oil vapor 9 Areas subject to salty air 10 Areas containing explosive gase...

Страница 19: ...is as follows Temperature 10 to 40 C 50 to 95 F Relative humidity 30 to 75 no condensation Air pressure 700 to 1060 hPa 4 Install in the area equipped with light fixture to ensure the product s opera...

Страница 20: ...n Manual 3 Grounding This product is grounded by plugging the medical grade plug into a medical grade outlet 4 Movement and Transport To move or transport this equipment contact the dealer where the e...

Страница 21: ...e Perform only the operations described in this manual Always following the instructions in the labels affixed to this product 1 1 Intended Use The Panoura18S dental panoramic and cephalometric device...

Страница 22: ...Before Using this Product 20 PANOURA 18S Operation Manual For a Cephalo attached equipment R E L K N M O ABCDEGHP Q ST...

Страница 23: ...Before Using this Product 21 PANOURA 18S Operation Manual...

Страница 24: ...Before Using this Product 22 PANOURA 18S Operation Manual...

Страница 25: ...sing this Product 23 PANOURA 18S Operation Manual 1 3 Names and functions of units and their parts 1 3 1 Main body 1 2 3 4 6 18 7 10 13 11 16 14 9 8 17 5 12 15 19 22 23 For a Cephalo attached equipmen...

Страница 26: ...st unit and moves up or down when the Up or Down switches are pressed Equipped with a positioning mirror on the front a cover and the power switch at the bottom 12 Control panel Enables the operator t...

Страница 27: ...TMJ and Single to select any of these four radiation modes The LED associated with selected mode will go on Panorama Single TMJ Child For a Cephalo attached equipment it toggles between Panorama Child...

Страница 28: ...ops sliding movement 8 Down switch When pressed and held down moves the sliding mechanism down to adjust the chin rest level When released stops sliding movement 9 Forward switch When pressed and held...

Страница 29: ...Displays CEPH on Tube voltage and Tube current indicator Tube current indicator Indicates the current setting for Tube current 4 Radiation time indicator Indicates the radiation time usually the rema...

Страница 30: ...diation mode goes off when it leaves the reset point Blinks while the arm is moving to the positioning beam radiation point goes off when it has reached the point 3 Cooling LED Keeps on during cooling...

Страница 31: ...the patient For automatic setting of irradiance condition see Irradiance condition for image acquisition set to AUTO P 29 1 For Cephalo image acquisition when image acquisition status is set to AUTO i...

Страница 32: ...align ing to the patient s reference face plane Frankfurt or Camper plane 3 Front tooth tomographic beam Used to adjust the patient s front tooth row and the X ray radiation point by aligning the bea...

Страница 33: ...nlocking position of the knob for detaching When locking When unlocking Caution Be sure to hold the handle when detaching the sensor The sensor may drop if you hold other parts Dropping the sensor may...

Страница 34: ...y absorbed data into electronic data by the signal from the control device and exports the image to the computer unit Uses Panorama Cephalo combined sensor 4 Docking station Cephalo For Cephalo image...

Страница 35: ...Before Using this Product 33 PANOURA 18S Operation Manual 1 3 9 Laser beam allocation Panoura 18S equips with a laser beam allocated in the position shown below Laser beam...

Страница 36: ...ng in contact with the patient have been made of the materials de scribed below 2 4 1 3 6 5 Name Function 1 Chin rest bloc ABS resin PC resin adaptor for edentulous jaw 2 Head support ABS resin 3 Hand...

Страница 37: ...rent flowing through the X ray tube X ray tube current 1 5 2 Operating principle This equipment is equipped with an X ray generator and a Sensor unit at an Arm unit sup ported by a Column unit and a s...

Страница 38: ...r as that in the panoramic mode the patient can be positioned within the field of photography to acquire only the images of left and right joint areas necessary for diagnosis Cephalometric radiography...

Страница 39: ...al radiation time about 16 seconds Child panoramic For child and small female patients the child panoramic mode can be used to acquire images in a shorter radiation time normal about 12 8 seconds TMJ...

Страница 40: ...or a Cephalo attached equipment 1 6 2 Selecting Radiation status You may select either radiation status of AUTO or MANUAL Radiation status AUTO The tube voltage tube current and radiation time are aut...

Страница 41: ...Before Using this Product 39 PANOURA 18S Operation Manual 1 6 3 Names of arm unit positions Start reset point Positioning point Patient introduction point Patient withdrawal point...

Страница 42: ...Before Using this Product 40 PANOURA 18S Operation Manual Cephalo radiography position Sensor detachment position For a Cephalo attached equipment 74 90...

Страница 43: ...equipment are as described in the following table X ray photographic cycles Panoramic mode Tube Voltage kV Tube Current mA Irr Time sec Cool Time sec Max radiations hr No of times Con 1 73 8 16 368 9...

Страница 44: ...MEMO...

Страница 45: ...to see if there exists nothing disturbing X ray photography Be sure to check the equipment in reference to 7 2 Regular checks P 112 and Regular check list P 112 prior to use Post operational procedure...

Страница 46: ...ol panel See also 6 1 2 Problems associated error messages displayed on the screen P 109 2 Pressing the Emergency Stop switch shuts off power supply to the main body causing X ray radiation and image...

Страница 47: ...er on usually the arm unit will move automatically to the patient introduction point Don t put anything within the movable range of the arm unit otherwise its sur rounding physical proprietaries may b...

Страница 48: ...mpatible with TWAIN Version 2 1 See also Regarding detailed operations of the viewer refer to the instruction manual of the viewer in use From the viewer in use select PANOURA18S as the source of imag...

Страница 49: ...he mode is switched when Mode switch is pressed Adult Panoramic Acquires panoramic images of the patient s tooth row in radiation time normally 16 sec Panorama Single TMJ Child Child Panoramic Acquire...

Страница 50: ...time and use the and switches to modify the settings To which mode is switched when Confirm switch is pressed Shifting to new setting for tube current Shifting to new setting for tube voltage Setting...

Страница 51: ...the tomographic position at 0 2 Click New in the main screen of Image Creator and Image Acquisition Condi tion Setting screen is displayed 3 In order to switch between Panoramic Acquisition Mode and...

Страница 52: ...ven through clicking Cancel Suppl The value set on either of the control panel and the PC is reflected Checking the chin rest block Check that the chin rest block appropriate for the selected radiatio...

Страница 53: ...the chin rest block base Use the TMJ adaptor CHILD for the child patient to acquire the highest qual ity of images of the temporomandibular joint Caution Be sure to use the adaptor appropriate for th...

Страница 54: ...of the equipment has been reset Otherwise the patient may be caught by the arm unit being injured Don t turn the arm unit manually otherwise the patient may be caught by the arm unit being injured or...

Страница 55: ...the equipment may be damaged If the sensor comes in contact with the patient s shoulder ask the patient to grasp the handle with his her arms crossed to lower the position of the shoulder Caution If n...

Страница 56: ...troduce the wheelchair into close to the column Caution The images of the patient on a standard wheel chair may be acquired The wheelchair with a longer backrest or custom made wheelchair may disturb...

Страница 57: ...y not be appropriately performed To avoid this problem be sure to return the arm to its reset point and then retry Aligning the median beam Once the patient has been introduced inside the equipment pr...

Страница 58: ...right sides Corrective measure Align the reference median line vertical line to the patient s midsagittal line 2 One half of the patient face enlarged and the other reduced Problems One half of the p...

Страница 59: ...nt may be injured Suppl The positioning beam automatically goes off 60 seconds after activation If the positioning beam has gone off before positioning has been successfully finished press the Beam sw...

Страница 60: ...f the middle point of the mandibular third tooth At the end of positioning check that all the beams have focused on their cor rect points The positioning switch once pressed turns on with its associat...

Страница 61: ...he horizontal reference line horizontal line 2 The patient s face inclined upward Problems Blur maxillary front teeth Patient ss hard palate overlapped on the root apexes on the maxillary tooth row Po...

Страница 62: ...of adjustment try to turn on the positioning beams to check that they focus on their correct positions Caution Don t force the head support open or close without gripping its base Otherwise the patien...

Страница 63: ...e Beam switch to activate the positioning beam Caution Press the correct Positioning switches such as the UP and Down switches in position ing Otherwise the patient may be injured Don t apply heavy lo...

Страница 64: ...he beam Aligning oculoauricular beam Align the horizontal line of the head support parallel to the camper plane on patient s head Moves the sliding mechanism up or down to adjust the chin rest height...

Страница 65: ...e head support open or close without gripping its base Otherwise the patient may be injured Turning off positioning beam Press the Beam switch to turn off the positioning beam Return the positioning m...

Страница 66: ...position ing prior to taking photographs Starting X ray radiation 1 After the operations described in 2 3 3 Setting the radiation mode P 47 make sure that Image Creator is displaying its image acquisi...

Страница 67: ...ate the controlled area while X ray is irradiated Photography in TMJ mode Caution To minimize the X ray dose irradiated on the patient be sure to perform correct position ing prior to taking photograp...

Страница 68: ...mode X ray radiation stops with X ray radiation LED off and an audible alarm silenced in any other area than temporomandibular joints Note that the X ray radiation switch must be pressed and held dow...

Страница 69: ...the equipment be sure to escort the patient while stepping back to the safe area Otherwise the patient may be injured by an accident e g top pling down 2 3 9 Image Transfer 1 Once the X ray irradiatio...

Страница 70: ...77 See also Regarding detailed opera tions of the viewer refer to the instruction manual of the viewer in use Caution Be sure to handle the hard disk correctly otherwise all the important data and pro...

Страница 71: ...t Turn the ear rod and nose support of the Cephalo Unit to the appropriate position according to the desired image acquisition mode namely Anteroposte rior Image PA or Lateral Image LA The arrow indic...

Страница 72: ...oach the equipment until movements of the arm unit to reset or to the cephalo image acquisition position are complete Otherwise the patient might have his her body caught in the arm unit and might cau...

Страница 73: ...quipment and cause an accident If an object is placed within the range of the arm unit s movement or underneath the cephalo unit it might cause damage to the object or malfunction to the equipment If...

Страница 74: ...a lateral LA image adjust the height of the nose support at the center between the patient s eyebrows and gently press it to fix Remove the nose support when acquiring an anteroposterior PA image or...

Страница 75: ...y pressed when positioning the patient might get injured Be cautious as to avoid erroneous operations Do not move the equipment up down or forward backward while the ear rods are in the patient s poru...

Страница 76: ...he finger is removed from the X ray radiation switch while X ray is being irradiated the X ray irradiation and image acquisition operation stop The image acquisition ends up in an error Re acquisition...

Страница 77: ...e equipment starts transferring the image data to the computer Image Creator Once the transfer is complete Image Creator starts formulating and display ing the cephalo image 2 Press OK to activate Ima...

Страница 78: ...tant data or programs inoperative Use a backup program etc to copy the contents of the hard disk drive into an auxiliary storage device Please note that we are not liable for the loss of data because...

Страница 79: ...l Imaging Software NOTE Prior to installing Image Creator check that PANOURA18S ControlManager has been installed If PANOURA18S ControlManager has not been installed install PANOURA18S ControlManager...

Страница 80: ...trolManager 78 PANOURA 18S Operation Manual 5 When the contents of the CD drive are displayed open ControlManager by double clicking it Select ControlMan ager 6 Click Open 7 Return to the Run dialog C...

Страница 81: ...al 3 1 2 Installing ControlManager 1 The right hand dialog will open 2 Then another right hand dialog will open Click Next 3 The License Agreement dialog will open Read the description of the software...

Страница 82: ...8S ControlManager 80 PANOURA 18S Operation Manual 4 The Location dialog will open Check the file location and click Next 5 The right hand dialog will open Click Install 6 During installation the right...

Страница 83: ...mputer is turned on PANOURA18S ControlManager is automatically activated in the background While PANOURA18S ControlManager is active is displayed on the task tray Configuring a communication port s 1...

Страница 84: ...e indicated 1 Tube voltage 2 Tube current 3 Radiation time 4 Radiation mode 5 Radiation status 3 4 Descriptions of Setup screen Software PANOURA18S ControlManager can be used to configure User Program...

Страница 85: ...e button associated with the setting to be modified 4 The right hand dialog will open To confirm the modification click OK Suppl In the User Program Mode any operation from the radiation switch or the...

Страница 86: ...isted as shown in the following screen 1 2 3 4 5 1 Main program version 2 Control panel program version 3 Hand Switch Box program version 4 Cephalo control relay board program version 5 PANOURA18S Con...

Страница 87: ...l 3 6 Configuring a communication port s 1 Click About in Main Menu 2 The Version Confirmation screen will open Click Port Setting 3 The Port Setup dialog will open 4 From the pulldown menu sele ct yo...

Страница 88: ...rial ports hereafter called ports has a designation assigned on the PC e g COM3 COM6 The assigned numbers are not always serial numbers 3 With more than one port incorporated in the equipment you need...

Страница 89: ...qualified person can install Image Creator NOTE Prior to installing Image Creator check that PANOURA18S ControlManager has been installed If PANOURA18S ControlManager has not been installed install PA...

Страница 90: ...S Operation Manual 4 The Browse dialog will open From Look in select the CD drive as a directory path 5 The files in Image Creator directory will be listed Select setup exe 6 Click Open 7 Return to th...

Страница 91: ...1 2 Installing Image Creator 1 The right hand dialog will open 2 Then another right hand dialog will open Click Next 3 The License Agreement dialog will open Read the description of the software licen...

Страница 92: ...age Creator 90 PANOURA 18S Operation Manual 4 The Location dialog will open Check the file location and click Next 5 The right hand dialog will open Click Install 6 During installation the right hand...

Страница 93: ...Directions for Use of Image Creator 91 PANOURA 18S Operation Manual 7 Once installation has finished another right hand dialog will open...

Страница 94: ...E Before activating Image Creator make sure that Dental Imaging software is installed In case Dental Imaging software is not installed follow the instructions below and install it 1 Insert the PANOURA...

Страница 95: ...of Image Creator 93 PANOURA 18S Operation Manual 5 Double click and open the Image Creator folder Select AJAT_DEN TAL_IMAGING exe 6 Press Open 7 Return to the Run dialog box Click OK The installer be...

Страница 96: ...ompting for installation destination Check the installation destination and click Next 3 A dialog box as in the figure on the right is displayed Select Full installation with hardware support and clic...

Страница 97: ...5 PANOURA 18S Operation Manual 4 A dialog box as in the figure on the right is displayed Click Next 5 A dialog box as in the figure on the right is displayed Click Next 6 A dialog box as in the figure...

Страница 98: ...ftware is automatically activated and takes up about 1GB of internal memory 4 3 Starting up Image Creator When the computer is restarted after the installation Image Creator is automatically activated...

Страница 99: ...Directions for Use of Image Creator 97 PANOURA 18S Operation Manual 4 5 Image Creator Screen Image Creator Main Screen without panoramic images Image Creator Main Screen with panoramic images...

Страница 100: ...e position of tomographic trajectory is se lected for the panoramic image displayed in the preview and the panoramic image to be newly created After the image acquisition the position of to mographic...

Страница 101: ...aded from the file Select and Export By pressing Export Button the acquired data can be named and saved By pressing Se lect Button a file from the saved data can be selected and loaded NOTE The file s...

Страница 102: ...ular domain on the panoramic image and pressing Auto Focus the panoramic image is reformulated by the selected domain being focused and the tomographic domain being automatically readjusted NOTE When...

Страница 103: ...tton is used when moving the image shown at the Image Display area horizontally Horizontal movemnet is done by the mouse s left dragging Zoom Button Zoom Button is used when enlarging or reducing the...

Страница 104: ...viewer For viewers capable of handing 16 bpp grayscale images the formulated image is transferred as it is For viewers only capable of handling 8 bpp grayscale images the image is transferred after th...

Страница 105: ...r or Panoura18S Control Manager software installed on the computer Refer to Descriptions of User Program Mode in the end of this section for the program number and display contents Operation through I...

Страница 106: ...8S Operation Manual 2 Click and select the User Program Number that needs adjustment 3 Press the button of the setting value needing adjustment 4 A dialog box as in the figure on the right is displaye...

Страница 107: ...Cervical spine correction Lo Off no correction Hi step 2 Lo step 1 Prog No 06 Specifies whether a service maintenance message appears once 2 000th radia tion has finished Service maintenance message...

Страница 108: ...MEMO...

Страница 109: ...ot face front Ask the patient to face front Whitish and blur front tooth area With the oblique cervical spine X ray is absorbed leading to a shadow formed on the front tooth area Ask the patient to ho...

Страница 110: ...to the main body or the cable has been correctly connect ed to the PC message has appeared The PC connection cable has not been correctly connected to the PC Connect correctly the cable to the PC The...

Страница 111: ...on the control panel to recover from the error If this error persists contact your dealer Er54 The arm unit movement does not finish within the given time period Er56 The back and forth driving motor...

Страница 112: ...ice maintenance services Total No of radiation has exceeded 2 000 It is recommended that you will perform regular checks Contact your dealer Fi02 The radiation switch has been pressed dur ing selectin...

Страница 113: ...r dampened soft cloth and then wipe away remaining moisture with a dry cloth 2 Cleaning plated surfaces Wipe out dirt stuck on the plated surfaces of the equipment with a water dampened soft cloth and...

Страница 114: ...pre operation 3 The unit has sit on the floor stably with no rattle For more information see Step 1 of 7 4 Performing check services P 114 pre operation 4 The unit moves up and down normally with no a...

Страница 115: ...actor The recommended main check services to be performed by the subcontractor are as described in the following table For more information the maintenance checking sheet supplied with the equipment L...

Страница 116: ...r abnor mal noise 4 Image acquisition 1 Acquire images in the Panoramic mode in the empty X ray room for 16 seconds 2 Check that the images have been imported in the PC Observe the machine behavior du...

Страница 117: ...rounded power outlet 1 Accuracy of a voltmeter tester AC Voltage 0 7 2 digits or its equivalents 2 Tester type Voltmeter for effective voltage measurement 3 Recommended testers 1 Fluke 87 V Fluke 2 PC...

Страница 118: ...MEMO...

Страница 119: ...household waste For proper treatment recovery and recycling please take these products to designated collection points where they will be accepted on a free of charge basis Alternatively in some count...

Страница 120: ...MEMO...

Страница 121: ...tient positioning part Patient positioning part list Item Name Quantity Chin rest block 1 Adaptor for edentulous jaw 1 TMJ adaptor ADULT 1 TMJ adaptor CHILD 1 Bite block consumable accessory 1 Caution...

Страница 122: ...MEMO...

Страница 123: ...uipment contains no time change component 10 2 Consumable stores The equipment contains consumable stores listed in the following table If you need to acquire any one of these consumable stores contac...

Страница 124: ...MEMO...

Страница 125: ...0 0 63 0 80 1 00 1 20 1 60 2 00 2 50 3 20 8 0 10 0 seconds Cephalo Carpus image acquisition Electric power supply resistance Maximum 0 2 or less Photographing direction Panorama TMJ Clockwise Counter...

Страница 126: ...sition base mount short type with an optional base 1 190kg Standing position wall mount with Cephalo 185kg Standing position wall mount short type with Cephalo 215kg Standing position base mount with...

Страница 127: ...ilter Specifications Aperture cover 0 2mmAl PC 1 0mm Acrylic filter 0 3mmAl Acryl 3 0mm Aluminum filter 1 5mmAl Aluminum 1 5mm X ray tube Focal spot Inherent filtration of X ray tube 0 4mmAl Inherent...

Страница 128: ...Main specification 126 PANOURA 18S Operation Manual 11 3 Dimensional drawing 11 3 1 Standing position wall mount Panorama type Unit mm...

Страница 129: ...Main specification 127 PANOURA 18S Operation Manual 11 3 2 Standing position wall mount short type Panorama type Unit mm...

Страница 130: ...Main specification 128 PANOURA 18S Operation Manual 11 3 3 Standing position wall mount long type Panorama type Unit mm...

Страница 131: ...Main specification 129 PANOURA 18S Operation Manual 11 3 4 Standing position base mount with an optional base Panorama type Unit mm...

Страница 132: ...Main specification 130 PANOURA 18S Operation Manual 11 3 5 Standing position base mount short type optional base Panorama type Unit mm...

Страница 133: ...Main specification 131 PANOURA 18S Operation Manual 11 3 6 Standing position base mount long type optional base Panorama type Unit mm...

Страница 134: ...Main specification 132 PANOURA 18S Operation Manual 11 3 7 Standing position base mount with an optional wide base Panorama type Unit mm...

Страница 135: ...Main specification 133 PANOURA 18S Operation Manual 11 3 8 Standing position base mount short type with an optional wide base Panorama type Unit mm...

Страница 136: ...Main specification 134 PANOURA 18S Operation Manual 11 3 9 Standing position base mount long type with an optional wide base Panorama type Unit mm...

Страница 137: ...Main specification 135 PANOURA 18S Operation Manual 11 3 10 Standing position wall mount Cephalo attached type Unit mm...

Страница 138: ...Main specification 136 PANOURA 18S Operation Manual 11 3 11 Standing position wall mount short type Cephalo attached type Unit mm...

Страница 139: ...Main specification 137 PANOURA 18S Operation Manual 11 3 12 Standing position wall mount long type Cephalo attached type Unit mm...

Страница 140: ...Main specification 138 PANOURA 18S Operation Manual 11 3 13 Standing position base mount with and optional base Cephalo attached type Unit mm...

Страница 141: ...Main specification 139 PANOURA 18S Operation Manual 11 3 14 Standing position base mount short type optional base Cephalo attached type Unit mm...

Страница 142: ...Main specification 140 PANOURA 18S Operation Manual 11 3 15 Standing position base mount long type optional base Cephalo attached type Unit mm...

Страница 143: ...Main specification 141 PANOURA 18S Operation Manual 11 3 16 Standing position base mount with an optional wide base Cephalo attached type Unit mm...

Страница 144: ...Main specification 142 PANOURA 18S Operation Manual 11 3 17 Standing position base mount short type with and optional wide base Cephalo attached type Unit mm...

Страница 145: ...Main specification 143 PANOURA 18S Operation Manual 11 3 18 Standing position base mount long type with and optional wide base Cephalo attached type Unit mm...

Страница 146: ...eld Double Insulated Wire Basic Insulated Wire etc X Ray Generator Pressor Circuit Head Thickness Sensor X Axis Positioning Sensor Arm Rotation Positioning Sensor Pulse Interface Unit Class 2 Laser Po...

Страница 147: ...ation Display Unit Operating Unit Main Control Circuit Lifting Switch Drive Circuit Drive Circuit Option Arm Rotation Stepping Motor X Axis Driving Stepping Motor Collimator Stepping Motor Hand Switch...

Страница 148: ...Main specification 146 PANOURA 18S Operation Manual 11 6 Spatial resolution MTF 0 2 4 6 0 2 0 4 0 6 0 8 1 Spatial Frequency LP mm MTF...

Страница 149: ...specification 147 PANOURA 18S Operation Manual 11 7 Anode cooling curve and heat content of X ray tube assembly 0 100 200 300 400 30 40 50 60 Time min Temp 0 100 200 300 400 0 100 200 300 Time min Hea...

Страница 150: ...MEMO...

Страница 151: ...Qualification Assurance QA Procedure 149 PANOURA 18S Operation Manual 12 Qualification Assurance QA Procedure 12 1 Structure of the phantom...

Страница 152: ...setting Image acquisition mode Standard panorama Image acquisition status Manual Tube voltage 58kV Tube current 2mA Radiation time 16sec User program mode setting Cervical spine correction OFF Image C...

Страница 153: ...Radius of the X direction should be within 19 of the radius of Y direction X direction Y direction See also For measurement functions see the viewer manual 2 Check the distance from the steel ball in...

Страница 154: ...MEMO...

Страница 155: ......

Страница 156: ...18S Operation Manual Issue date xxxx 2010 1 st version European Authorized Representative Emergo Europe Molenstraat 15 2513 BH The Hague The Netherlands Manufactured by The Yoshida Dental MFG Co Ltd 1...