<9. TROUBLESHOOTING>
9-3
IM 12E01B30-02EN 2nd Edition : June 09 , 2021-00
9.2
When No Error Indication Appears
There is a possibility that measurement has abnormalities without any error occurrence. Next
table shows some possible causes. Solve the problems referring to the next table.
Table 9.3
Abnormalities, Their Possible Causes, and Remedies
Abnormalities
Possible Causes
Remedies
Measured value
fluctuates largely.
● Sample air bubble or contamination
stays in flow cell.
● The measured value fluctuates
according to the sample turbidity.
●
Clean the Flow cell..
●
Bring the air denoising head tank which
is equipped upstream of the sample
inlet, to a normal state.
●
Adjust the sample outlet valve to
increase the pressure inside the Flow
cell.
●
Check if the turbidity reading is stable
by using Zero water or Check tool. If
the reading is not stable, identify where
the error occurs in Flow cell, Light
source, or Detector board. After then,
bring it to a normal state.
●
If no error is found in the instrument
or sample flow, set larger value of
measurement time constant.
Measured value
increased over time
● Inside of the detector cell is
contaminated.
● Air bubbles adhere to the surface of
glass window.
● Glass window has condensation.
● Glass window is scratched.
●
Clean the glass window or inside wall of
the Flow cell.
● Reduce sample flow rate by controlling
the valve at the outlet to higher the
pressure level inside the Flow cell.
●
Check the condition of desiccant.
Replace it if necessary with new one.
●
If the measured value does not go down
even after trying the remedies from 1
to 3 above, check if the glass window
is scratched. If any scratch is found,
replace the Flow cell.
Measured value
abruptly increases
but plunged in a
short time
● Air bubbles or contamination pass
through the flow cell.
● Reduce sample flow rate by controlling
the valve at the outlet to higher the
pressure level inside the Flow cell.
●
Turn on the Air denoising function.
●
Optimize the air denoising parameter.
Measured value
does not vary.
● Light source is abnormal.
● Communication between Detector
and Liquid analyzer is abnormal.
● Check the light source is lit on.
● Check the cable connection between
FLXA402T Analyzer and TB820D
detector.