<8. MAINTENANCE>
8-3
IM 12B01B30-02EN 2nd Edition : June 09 , 2021-00
(8) Supply zero water to the detector. When the Flow cell is filled with the water, drain it. Repeat
this at least three times to thoroughly rinse off detergent in the Flow cell. See step (3) for
valve operation.
(9) Replace the top cover.
(10) To go to measurement mode, turn off the maintenance mode switch.
(11) Washing is now complete.
8.2
Washing the Head Tank
This section describes the procedure for washing the head tank.
(1) Turn on the maintenance mode.
(2) Close the sample supply valve.
(3) Draining water from the head tank and washing it.
The following describes how to wash two optional head tanks provided by Yokogawa.
[Pressurized head tank]
1. Close the sample valve.
2. Loose and remove the drain plug attached to the side of the head tank, to drain the
sample.
3. Remove the cover on the top of the head tank. Wash inside the tank using a sponge or
soft cloth soaked in mild detergent. Pour tap water into the tank from above and rinse off
thoroughly.
4. Replace the drain plug and the top cover.
5. Open the sample valve to run the sample into the head tank and wait until the tank is
refilled.
[Simple head tank]
With the sample valve open, open and close the drain valve under the head tank at least three
times to rinse out the head tank.
(4) After washing the head tank, open the sample supply valve to supply the sample to the
detector.
(5) To go to measurement mode, turn off the maintenance mode.
(6) Washing is now complete.