<4. SETTING PARAMETERS>
4-6
IM 12E01B30-02EN 2nd Edition : June 09 , 2021-00
4.2
Cal./Maint. Settings
4.2.1
Turbidity value of check tool
Displays the value on the turbidity check tool to measure the slope. The value is set at the factory.
4.2.2
Zero shift value, Correction factor
These parameters are related to zero shift correction and sensitivity correction in the calibration
menu. Normally zero shift value is 0, and correction factor is 1. See chapter 7. Calibration.
4.2.3
Stability check
You can configure the following parameters to evaluate the measurement stability in a calibration.
● Stability check allowable width
● Stability checking time
● Stability checking time
The measurement stability is verified, when the measurement stays within the range of “Stability
check allowable width” without exceeding the range for longer than the set period of” Stability
checking time”.
If the measurement does not become stable before “Stability check limit time” is up, a
confirmation dialog appears to abort the calibration.
l
Stability check allowable width
Set the range over which the stability of a measured value is checked. If variations of a measured
value over the stabilization time are within this setpoint, the measured value is judged to have
stabilized.
l
Stability checking time
During calibration, the stability of the measured value is constantly monitored. If variations of
a measured value over the stabilization time are within the Stability check allowable width set
above, the measured value is judged to have stabilized.
l
Stability check limit time
If the measured value does not become stable during the period of Stability check limit time, the
calibration is cancelled.
4.2.4
Wash/Cal. box
Select whether or not to use Wash Box (Relay box for solenoid valve). While using Relay box,
make Wash box Enabled.
If Wash Box is disabled, you cannot implement Auto wash/calibration.
If you enable this function when Wash Box is not connected, an alarm occurs to notice Wash Box
error.
4.2.5
Ultrasonic wash
Select Ultrasonic wash On/Off. When this function is On, you can switch Ultrasonic wash On/Off
on Maintenance menu. If you need to stop temporarily, turn it Off on the Maintenance menu.