<3. PREPARATION for OPERATION>
3-4
IM 12E01B30-02EN 2nd Edition : June 09 , 2021-00
3.8
Adjusting ultrasonic power
When “-U1” for the Oscillator for ultrasonic cleaning is specified, adjust the ultrasonic power.
Be sure to fill the flow cell with a sample when adjusting the ultrasonic power. First, turn on the
ultrasonic power from the maintenance menu of FLXA402T.
Then, perform the adjustment of the PG400 pulse generator for a clean unit.
(1) Open the PG400 cover and check the output switch. If it is OFF, turn it ON.
(2) Confirm that the green POWER LED and red OUTPUT LED are lit on.
(3) Adjust the oscillator output so that it does not affect the turbidity readings. Adjust the
output volume of the PG400 so that the cavitation effect of ultrasonic cleaning affects
the turbidity indication within 1% of the output range (full scale) used. The volume of
the output range is set to the middle at the factory setting.
For details on the adjustment method, see the PG400 user’s manual
.
3.9
Switching to measurement mode
After supplying the sample and adjusting its flow rate, activate all the instruments in the loop
configuration, such as ultrasonic cleaning equipment. When it is ready to operate, switch off the
maintenance mode to shift to the measurement mode. Make sure no error message appears.