<8. MAINTENANCE>
8-9
IM 12B01B30-02EN 2nd Edition : June 09 , 2021-00
Fuse
Figure 8.8
Fuse Replacement
n
Relay box fuse replacement (when Suffix code -A5 is selected)
Be sure to use A1624EF for the fuse of the Relay box.
(1) Loosen the four screws and open the Relay box cover.
(2) Remove the two fuse holders. (See Figure 8.9)
(3) Replace the fuse with a new one and attach the fuse holder.
(4) Close the relay box cover.
Fuse
Figure 8.9
Fuse Replacement (Relay box)
CAUTION
After turning on the power of the instrument, warm it up for at least 30 minutes before starting the
measurement.
Arrange the fuse as below.
Table 8.5
Fuse
Part No.
Name
Description
A1633EF
FUSE
1 fuse for TB820D
A1624EF
FUSE
1 fuse for Relay box (*)
(*)
2 fuses are required for Relay box.