<8. MAINTENANCE>
8-8
IM 12B01B30-02EN 2nd Edition : June 09 , 2021-00
Fix the lock lever.
When properly fixed,
it clicks.
Plug in the connector.
The side with the solder part faces up.
<Right>
<Wrong>
Pinch the strap and push down
the lock lever.
Figure 8.7
Connecting cables
8.7
Fuse Replacement
If the fuse blows, replace the fuse.
There is one fuse in the FLXA402T, one in the TB820D, and two in the Relay box (when Suffix
code -A5 is selected). Read IM 12A01G01-02JA for fuse replacement of FLXA402T.
Replace the fuses of TB820D and Relay box as follows.
CAUTION
Be sure to turn off the power of the instrument before starting the fuse replacement.
CAUTION
Before opening the front cover of the detector, turn off the power to the TB820D turbidimeter.
Make sure that the turbidimeter is not powered before work. Do not touch terminals while the
turbidimeter is being powered.
n
TB820D fuse replacement
Be sure to use A1633EF for the fuse of TB820D.
(1)
Loosen the four (4) setscrews and remove the Smart unit cover.
(2)
Take the fuse holder out.
(3)
Replace the fuse with a new one.
(4)
Put the Smart unit cover in place.