< 7. Calibration >
7-10
IM 12E01B30-02EN 2nd Edition : June 09 , 2021-00
7.9
2 points correction
The Zero point and Sensitivity correction factor are adjusted. Use two arbitrary known turbidity
solution. Take two measurements and record the low concentration as low point, the high
concentration as high point. Be sure to take low point first, then do the measurement of the high
point.
[How to perform a two point calibration]
1.
Turn on the maintenance mode.
2
Feed the solution for Low point measurement into the Flow cell.
3 Press
on Main screen to go to the calibration.
4
Select [2point calibration] in the menu list of Calibration others.
5
Confirm the reading stable. Press [check stability].
6
When a message pops saying that the reading is now stable, enter the calibration value.
Press [Next].
7
Feed the solution for High point measurement into the Flow cell. Wait until the reading
becomes stable. After it becomes stable, press [check stability].
8
When a message pops saying the reading is now stable, enter the calibration value.
Press [Next].
9
Confirm the result of calibration. Press [Accept] to update the parameter for Zero shift
value and sensitivity correction factor.
10
Drain the sample and clean the Flow cell with zero turbidity water. Place the clear cover
and the top cover on top of the detector. If you finish the calibration, run the sample and
turn off the maintenance mode.