<3. PREPARATION for OPERATION>
3-1
IM 12E01B30-02EN 2nd Edition : June 09 , 2021-00
3. PREPARATION for OPERATION
n
Operation mode
Two types as follows are available.
l
Measurement mode
Steady-state operation takes place.
l
Maintenance mode
Perform maintenance in this mode. Be sure to turn to this mode before starting calibration or
maintenance.
To switch to Maintenance mode, Main screen > Maintenance mode OFF > ON
(or Sensor menu > Maintenance > Maintenance mode)
n
Operation in Maintenance mode
Item
Operation
Measurement
Maint. damping time constant applies
mA output
Hold or output by following the automatic hold setting.
Converter menu > Converter setting> mA output setting >
Output channels > Auto hold setting
Contact output
When maintenance is setup, the contact is On.
Other contact outputs remain the previous value gained right
before the Maintenance turned on.
Manual
operation
Solenoid manual operation is enabled.
Sensor menu > Maintenance > Valve. Pump
To initialize Valve, turn off the maintenance mode.
Auto wash Auto
calibration
not available
3.1
Piping and Wiring Check
Confirm that the piping and wiring have been completed correctly before supplying water or power.
Refer to the chapter 2 for Installation and piping.
3.2
Confirming detector condition
Ensure that the drain cock is firmly closed. Open the front cover and set a new desiccant. Close
the cover firmly.
3.3
Supplying Power and switching operation mode
Confirm that the piping and wiring are made correctly. Turn on the power supply then the detector
starts up in a measurement mode. Keep On the Maintenance Mode until the preparation for the
operation completes. Turn on
on Main screen of the converter to go to Maintenance mode.
when the “-U1” (Oscillator for ultrasonic cleaning) is selected, the ultrasonic cleaning is ON when
the instrument starts up. Keep the ultrasonic OFF until a sample is introduced.
Sensor menu >
Maintenance > Ultrasonic On/Off