<4. SETTING PARAMETERS>
4-8
IM 12E01B30-02EN 2nd Edition : June 09 , 2021-00
4.5
Auto Wash/Calibration setting
4.5.1
Auto sequence for wash/cal.
Enable this function to use Auto wash/calibration. If this function is disabled, Auto wash/
calibration is never implemented even if other setting is active.
4.5.2
Start date
As the first time Auto wash/calibration, enter a future date/time but not the current one.
If you want to perform the next and subsequent Auto wash/calibration automatically, turn ON Auto
update of start date. The next date of Auto wash/calibration is determined according to washing
interval.
4.5.3
Auto Wash/Calibration function and wash type
You can set Auto wash function and Auto calibration function respectively to Enable/Disable.
When you use Auto calibration, always set Auto wash to Enable. Always select water flow as
wash type.
4.5.4
Auto sequence setting
Configure the following parameters to setup the schedule for Auto wash and Auto calibration.
● Washing interval
● Repeat count of wash
● Calibration frequency
The auto-sequence starts at the setup hour of the start time at the washing interval cycle. Set
how many times the auto wash repeats in a sequence by setting “Repeat count of wash”.
When Auto calibration function” is ON, Auto sequence starts at an interval determined at
Calibration frequency. When Auto wash ends, Auto calibration operates one time.
When Calibration frequency is set to “0”, no Auto calibration operates.
The next figure shows the example when:
Repeat count of wash is 2.
Calibration frequency is 3.
Automatic calibration takes place one time after the auto sequence repeats as many as
Calibration frequency.
start date
and time
Next sequence
start
Next sequence
start
Time schedule
Washing interval
Repeat count of wash
Washing interval
Repeat count of wash
Washing interval
Repeat count of wash
Wash Wash
Wash Wash
Wash Wash Cal
Wash
Wash
Other parameters are for setting details about one count of Auto wash/Auto calibration. For
further information on the auto sequence, read Appendix.