5-8
IM 720120-01E
Note
Be sure to hold the SL1000 firmly while pressing down on
the top section of the screwless
terminal
with a flat-blade screwdriver.
Use the following signal wires for the remote input.
• Signal wire thickness (single wire):
ϕ0.4 to 1.0 (AWG26 to 18)
(stranded wire):
0.3 to 0.75 mm
2
(AWG22 to 20)
• Recommended length of stripped wire: 10 mm
Circuit Diagram of the Remote Input
VHC14 or equivalent
120
Ω
+5 V
4.7 k
Ω
100 pF
REMOTE
5.5 Connecting the Remote Input Terminals (REMOTE)