3-16
IM 720120-01E
• For the 701251 (HS1M16)
Maximum input voltage (at a frequency of 1 kHz or less)
• In combination with an isolated probe (10:1) 700929 or (100:1) 701947, or a
passive probe (10:1) 702902
1
600 V (DC + ACpeak)
• Safety cable (1:1) (combined with 701954)
3
140 V (DC + ACpeak)
• Direct input (cable that does not comply with the safety standards)
5
42V (DC + ACpeak)
Maximum rated voltage to earth (at a frequency of 1 kHz or less)
• In combination with an isolated probe (10:1) 700929 or (100:1) 701947, or a
passive probe (10:1) 702902
2
Or safety cable (1:1) (combined with 701954)
4
400 Vrms (Measurement Category Other (O)), 300 Vrms (CAT II)
• Direct input (cable that does not comply with the safety standards)
6
42V (DC + ACpeak) (CAT II, 30 Vrms)
• For the 701255 (NONISO_10M12)
This module is non-isolated. Be sure to fasten the module screws when
measuring a voltage above 42 V on this module. In addition, use the dedicated
701940 non-isolated passive probe (10:1).
Maximum input voltage (at a frequency of 1 kHz or less)
• Combined with the 701940 passive probe (10:1)
600 V (DC + ACpeak)
• Direct input(cable that does not comply with the safety standards)
5
200 V (DC + ACpeak) (as a value that meets the safety standard)
250 V (DC + ACpeak) ( maximum allowable voltage, as a value that does not
damage the instrument when applied)
• For the 701275 (ACCL/VOLT)
Maximum input voltage (at a frequency of 1 kHz or less)
• Combined with the 701940 passive probe (10:1)
7
or direct input(cable that
does not comply with the safety standards)
5
42 V (DC + ACpeak)
Maximum rated voltage to earth (at a frequency of 1 kHz or less)
• Combined with the 701940 passive probe (10:1)
8
or direct input (cable that
does not comply with the safety standards)
6
42 V (DC + ACpeak) (CAT II, 30 Vrms)
• For the 701281 (FREQ) and 720281 (FREQ)
Maximum input voltage (at a frequency of 1 kHz or less)
• In combination with an isolated probe (10:1) 700929 or (100:1) 701947, or a
passive probe (10:1) 702902
1
420 V (DC + ACpeak)
• Safety cable (1:1) (combined with 701954)
3
or direct input (cable that
does not comply with the safety standards)
5
42 V (DC + ACpeak)
Maximum rated voltage to earth (at a frequency of 1 kHz or less)
• In combination with an isolated probe (10:1) 700929 or (100:1) 701947, or a
passive probe (10:1) 702902
2
701281: 300 Vrms (CAT II)
720281: 400 Vrms (CAT II)
• Safety cable (1:1) (combined with 701954)
4
or direct input(cable that
does not comply with the safety standards)
6
42 V (DC + ACpeak) (CAT II, 30 Vrms)
3.5 Connecting the Probes