Connecting a Current Probe
If you are using a YOKOGAWA’s current probe
*
, use the probe power supply (option) on
the front panel or a probe power supply sold separately (701934) to supply power to the
probe.
* YOKOGAWA current probes: the 701930, 701931, 701932, 701933,701917, and
701918
For details on the connection procedure, see the manual that comes with each probe.
PROBE P
OW
ER
( 12 V
DC
)
SLOT
1
SLOT
2
3
SLOT
SLOT
4
5
SLOT
SLOT
6
7
SLOT
SLOT
8
HIGH
-SPEED
DAT
A ACQU
ISITI
ON U
NIT
EXT CLK I
N
POWER
TRIG'
D
START
/
STOP
DISP
LAY
CAUTION
Use the probe power supply (option) on the front panel of the SL1000 only to
supply power to the current probes. Also, be sure to use only the number of probes
allowed. Otherwise, the SL1000 or the device connected to the probe power supply
terminal may break.
French
ATTENTION
Utiliser l’alimentation de la sonde (option) du panneau frontal de SL1000
uniquement pour alimenter les sondes de courant. De plus, s’assurer que seul le
nombre de sondes autorisé est utilisé. Le dispositif SL1000 ou l’appareil relié à la
borne d’alimentation de la sonde risquerait sinon de casser.
Precautions to Be Taken When Using the Current Probes
When connecting the current probe to the probe power supply terminal (option) on
the front panel, make sure that the current does not exceed the range shown below.
Otherwise, the SL1000 operation may become unstable due to the activation of the
excessive current protection circuit of the power supply.
PROBE POWER( 12 V DC )
A B C D
When each terminal is assigned names A through D
(Total current consumption by A through D) < 1300 mA
3-22
IM 720120-01E
3.5 Connecting the Probes