5-3
IM 720120-01E
External I/O T
erminals
1
2
3
4
5
6
7
App
Index
5.3 Connecting the Trigger Output Terminal (TRIG
OUT)
CAUTION
Do not apply external voltage to the TRIG OUT terminal. If you do, the SL1000 may
malfunction.
French
ATTENTION
Ne pas court-circuiter la borne TRIG OUT et ne pas appliquer de tension de sortie.
Cela pourrait endommager l’instrument.
Trigger Output Terminal
This terminal outputs a CMOS signal when a trigger occurs. The signal level is normally
high and switches to low when a trigger is activated.
TRIG OUT
Item
Specifications
Connector type
BNC
Output level
CMOS level (0 to 5 V)
Logic
Falls when the trigger is activated, rises after completing acquisition
Output delay
Within (100 ns + 1 sample period)
Output hold time
100 ns or longer
Circuit Diagram and Timing Chart of the Trigger Output
TRIG OUT
120
Ω
HC245 or equivalent
Trigger occurrence (time of trigger position)
Trigger output
Internal trigger
Output hold time
Output delay