1-4
IM 720120-01E
1.3 Display
Screens
The available SL1000 screens are the module status screen, error screen, and
communication parameter screen.
Module status
screen
Communication
parameter
screen
Error screen
DISPLAY
Startup or timeout
key
key
key
DISPLAY
DISPLAY
• At startup, the module status screen appears.
• Press DISPLAY to switch screens.
• If there is no key operation for 30 seconds, the screen returns to the module status
screen.
Screen Description
01:02
701
250
701
251
701
255
701
260
720
210
1 2 3 4 5 6 7 8
Group ID:unit ID
Module status icon
Status icon
IP address
Module status screen
Group ID:unit ID
Error information
Displays error, message code,
and time of occurrence
Status icon
Error screen
Group ID:unit ID
Status icon
Communication parameter screen
E: 913 09:31:20
E: 703 10:52:36
M: 052 13:02:27
M: 053 13:05:15
NO : TEMP01
IP : 192.168.0.100
01:02
01:02
Instrument number
(marked on the name plate. See page iii.)
2008/08/24 09:32:24
Date and time