<9. Connection Device Type>
9-17
IM 01W03D02-01EN
Object ID Attribute
ID
Label
Description
Default value Handling
*
1
8.
SENCOM
Interface
block
(continued)
96
Input 2
Impedance
Indicates impedance value of the Input 2.
This means the electrical resistance of the
reference electrode liquid junction, and used to
determine smudge of the reference electrode.
Input 2
Impedance
Status
0 kΩ
Input 2 Impedance Too Low
>= 2000 kΩ
Input 2 Impedance Too High
0.0
R
137
Stable Time
Indicates the time until the reliable data is available.
The unit is second.
0
R
20.
AI1
block
21.
AI2
block
22.
AI3
block
23.
AI4
block
1
Process Value
AI1, AI2, AI3, and AI4 are output object.
1. Value: output value of the object.
2. Status: indicates the status of the object’s
output value.
1. ---
2. ---
R
2
Block Mode
Select the block’s operation status. O/S and Auto
can be selected.
1. Target: Specify object mode of the object.
2. Actual: Indicates current mode of the object.
3. Permitted: Indicates the mode selected by
Target of the object.
4. Normal: Indicate normal status mode of the
object.
1.Auto
2. Auto
3. Auto
4. Auto
W(P)
3
Concentrator
OID
Indicates the Concentrator object value that
corresponds to the data update of the Process
Value.
---
R
4
Scale
Allows specifying the upper or lower limit for the PV
scaling, unit code, etc.
1. EU at 100%: Indicate the upper limit of the
PV value
2. EU at 0%: Indicate the lower limit of the PV
value
3. Units Index: Indicate the setting unit used for
the PV value
4. Decimal: Indicate the digit number below
the decimal point displayed in the integral
indicator
1. 100
2. 0
3. %
4. 2
W(P)
102
Tag Description A universal parameter to store the comment that
describes the tag.
AI1: AI1
AI2: AI2
AI3: AI3
AI4: AI4
W(P)
103
Simulate Switch A simulation function switch for the object.
1: Disable (default)
2: Enable
Disable
W(P)
104
Simulate Value
When Simulate Switch is set to Enable, this value
is used as the input value for the object. The input
value can be changed.
0
W(P)
106
PV Energy Left
Enable
Allows assign the Energy Left to AI1.PV.Value when
Sensor Type in TRANSDUCER block is set to “Not
Used”.
0: Disable (default)
1: Enable
0
W(P)
*1: R: Read only, W: Read and Write, (P): Target of the write protection
*2: NAMUR NE 107 “Self-Monitoring and Diagnosis of Field Devices”