<6. Operation>
6-4
IM 01C25W05-01EN
6.5.1 Draining Condensate from C10FR
Flushing Connection Ring
1. Gradually open the drain screw to drain from
the flushing connection ring.
2. When the flushing connection ring is completely
drained, close the drain screw. Tighten the drain
screw to a torque of 10 N·m {1 kgf·m}.
WARNING
If the process fluid is toxic or harmful, take
appropriate care to avoid contact with the body.
Drain screw
When you loosen the drain screw,
the accumulated liquid (or gas)
will be expelled in the direction
of the arrow.
F0605.ai
Figure 6.5
Draining Condensate from C10FR
Flushing Connection Ring
6.5.2 Venting Gas from C10FR Flushing
Connection Ring
1. Gradually open the vent screw to vent gas from
the flushing connection ring.
2. When the flushing connection ring is completely
vented, close the vent screw. Tighten the vent
screw to a torque of 10 N·m {1 kgf·m}.
WARNING
If the process fluid is toxic or harmful, take
appropriate care to avoid contact with the body.
Vent screw
When you loosen the vent screw,
the accumulated liquid(or drain)
will be expelled in the direction
of the arrow.
F0606.ai
Figure 6.6 Venting Gas from C10FR Flushing
Connection Ring