YOKOGAWA DPharp EJXC40A Скачать руководство пользователя страница 1

User’s

Manual

Digital Remote Sensor

IM 01C25W05-01EN

IM 01C25W05-01EN

4th Edition

Содержание DPharp EJXC40A

Страница 1: ...User s Manual Digital Remote Sensor IM 01C25W05 01EN IM 01C25W05 01EN 4th Edition...

Страница 2: ...adio Transceivers 2 3 2 8 Insulation Resistance and Dielectric Strength Test 2 3 2 9 Installation of an Explosion Protected Instrument 2 4 2 9 1 FM Approval 2 4 2 9 2 ATEX Certification 2 9 2 9 3 IECE...

Страница 3: ...1 Preparation for Starting Operation 6 1 6 2 Zero Point Adjustment 6 2 6 3 Starting Operation 6 3 6 4 Shutting Down Operation 6 3 6 5 Venting or Draining Transmitter Process Detector Section 6 3 6 5...

Страница 4: ...ail Setup 7 8 7 2 3 1 Pressure Signal Setup 7 8 7 2 3 2 Pressure Damping Time Constant 7 8 7 2 3 3 Impulse Line Connection Orientation Setup 7 9 7 2 3 4 Bi directional Flow Measurement Setup 7 9 7 2 3...

Страница 5: ...hecking the Blockage Detection Operating Capacity 7 42 7 3 2 2 9 ILBD Execution 7 43 7 3 2 2 10 Tuning 7 43 7 3 2 2 11 Resetting the Reference Value 7 45 7 3 2 2 12 ILBD Parameter List 7 45 7 3 2 3 He...

Страница 6: ...3 Setup A1 1 A1 2 4 Required Parameter Settings A1 1 A1 2 5 Proof Testing A1 1 A1 2 6 Repair and Replacement A1 2 A1 2 7 Startup Time A1 2 A1 2 8 Firmware Update A1 2 A1 2 9 Reliability Data A1 2 A1 2...

Страница 7: ...l and the transmitter body section style code written on the product nameplate F0101 ai DRS cable High pressure side transmitter master Low pressure side transmitter slave Screw mounting Digital remot...

Страница 8: ...NOTE When describing the model name like EJ 530 it shows the applicability for EJX530A Regarding This Manual This manual and the identification tag attached on the packing box are essential parts of...

Страница 9: ...t heeded the protection provided by this instrument may be impaired In this case Yokogawa cannot guarantee that the instrument can be safely operated Please pay special attention to the following poin...

Страница 10: ...e time of purchase Problems occurring during the warranty period shall basically be repaired free of charge If any problems are experienced with this instrument the customer should contact the Yokogaw...

Страница 11: ...U bolt L U bolt S U bolt S nut Figure 2 1 Transmitter Mounting Hardware 2 1 Model and Specifications Check The model name and specifications are written on the name plate attached to the case Look up...

Страница 12: ...thermal insulation and or ventilation Also avoid locations subject to prolonged high temperature and high humidity Ambient Atmosphere Do not install the transmitter in a corrosive atmosphere If this c...

Страница 13: ...erminals in the terminal box 4 Turn OFF the insulation tester Then connect the insulation tester plus lead wire to the shorted SUPPLY terminals and the minus leadwire to the grounding terminal 5 Turn...

Страница 14: ...setting switch must not be used in a hazardous area IMPORTANT For combined approval types Once a device of multiple approval type is installed it should not be re installed using any other approval t...

Страница 15: ...n installing the equipment In case Nonincendive Field Wiring Concept is used for the interconnection FM approved Associated Nonincendive Field Wiring Apparatus which meets the following conditions mus...

Страница 16: ...2 Handling Cautions 2 6 IM 01C25W05 01EN Note 6 Control Drawing...

Страница 17: ...2 Handling Cautions 2 7 IM 01C25W05 01EN...

Страница 18: ...2 Handling Cautions 2 8 IM 01C25W05 01EN...

Страница 19: ...vices in a hazardous location Note 4 Maintenance and Repair The instrument modification or parts replacement by other than authorized representative of Yokogawa Electric Corporation is prohibited and...

Страница 20: ...ment Note 5 Specific Conditions of Use WARNING Precautions shall be taken to minimize the risk from electrostatic discharge of painted parts When the enclosure of the Pressure Transmitters is made of...

Страница 21: ...nd Repair WARNING The instrument modification or parts replacement by other than an authorized Representative of Yokogawa Electric Corporation is prohibited and will void the ATEX certification Note 6...

Страница 22: ...5 female ANSI 1 2 NPT female M N or W 3 Installation WARNING All wiring shall comply with local installation requirements and the local electrical code 4 Operation WARNING OPEN CIRCUIT BEFORE REMOVING...

Страница 23: ...in the event of rare incidents an ignition source due to impact and or friction sparks is excluded Model EJX P and EJA P series pressure transmitters are not capable of withstanding the dielectric str...

Страница 24: ...Certification Electrical Connection A mark indicating the electrical connection type is stamped near the electrical connection port These marks are as followed F0206 ai Location of the mark Screw Siz...

Страница 25: ...X PROCESS TEMP 120 C Supply Output Ui 30V Ii 200mA Pi 0 9W Ci 27 6nF Li 0mH Communication Uo 8 2V Io 160mA Po 0 3W Co 7 6 F Lo 1mH POTENTIAL ELECTROSTATIC CHARGING HAZARD SEE USER S MANUAL KS24 S No F...

Страница 26: ...Note 2 Set the switches BO WR as shown in the figure above to set the burn out direction for CPU errors and write protection The Burnout switch is set to the H side for delivery unless option code C1...

Страница 27: ...cription 3 2 IM 01C25W05 01EN 3 2 Low Pressure Side Transmitter Slave F0302 ai Transmitter CPU assembly DRS cable connection port Amplifier cover Figure 3 2 Component Names of the Low Pressure Side Tr...

Страница 28: ...lead to leaks Zero adjustment screw Pipe back side of the instrument F0401 ai Figure 4 1 Pipe and Zero Adjustment Screw 4 2 Mounting IMPORTANT As the pressure detector diaphragm is thin and prone to d...

Страница 29: ...eting IMPORTANT Tighten the hexagonal nut part of the capsule assembly See Figure 4 4 Pressure detector section F0404 ai Figure 4 4 Tightening Transmitter F0405 ai Gasket Stud bolt Nut Figure 4 5 Moun...

Страница 30: ...e Figure 4 9 Correctly install the diaphragm seal part on the tank checking the label PRESSURE HIGH or PRESSURE LOW on each seal Exercise care so as not to damage diaphragm surfaces For models with di...

Страница 31: ...you connect the power supply directly it may damage the equipment 4 4 Mounting the C10FR Flushing Connection Ring 4 4 1 Mounting to Pressure Detector Section The C10FR flushing connection ring is moun...

Страница 32: ...re mounting the ring use the new spiral gasket as shown in below table Table 4 1 Spiral Gasket for Pressure Detector Section Side Hoop material Part number Gasket size Description SUS316 SUS316L Hast...

Страница 33: ...rt No F9145YN Gasket contact surface F0412 ai Diaphragm seal section Figure 4 14 Affixing the Teflon Film 4 6 Tank Spud Mounting Method A tank spud is used to joint a tank and an extended diaphragm It...

Страница 34: ...uck to the surface Special Tools for Tank Spud Welding Yokogawa provides special tools exclusively designed for the tank spud welding to make the installation easily and effectively These tools can be...

Страница 35: ...tscrews securing the transmitter section to the capsule assembly 2 Rotate the transmitter section slowly to the desired position 3 Tighten the two setscrews to a torque of 1 5 N m 15 kgf cm IMPORTANT...

Страница 36: ...e shielded wires in areas that are susceptible to electrical noise In areas with higher or lower ambient temperatures use appropriate wires or cables In environment where oils solvents corrosive gases...

Страница 37: ...a two wire transmission system signal wiring is also used as power wiring DC power is required for the transmitter loop The transmitter and distributor are connected as shown below For details of the...

Страница 38: ...an be run through a cable gland or a wiring metal conduit In either method follow the procedure below to connect and disconnect wiring CAUTION In the case of the flameproof type running the wires thro...

Страница 39: ...2NPT Type M20 Type Gasket Entry with Seal Running Coupler Backnut Entry with Seal Running Coupler Backnut F0511 ai Figure 5 11 Components of the Cable Gland Installing the cable gland 1 Disassemble t...

Страница 40: ...ods side opposite side of DRS connector before laying the cable Procedure 1 Disassemble the cable gland loosen the all parts 2 Remove the protection cap on the DRS electrical connection and DRS connec...

Страница 41: ...ugh the wiring metal conduit and screw the conduit into the wiring port CAUTION Do not pull on the cable or apply excessive mechanical shock to it 5 5 3 Removing DRS Cable with Cable Gland 1 Slowly re...

Страница 42: ...tion of transmitters Follow the domestic electrical requirements as regulated in each country Ground terminals are located on the inside and outside of the terminal box Either of these terminals may b...

Страница 43: ...roperly Check parameter values or change the setpoints as necessary See chapter 7 for the operating procedure of the HART configuration tool If the transmitter is equipped with an integral indicator i...

Страница 44: ...d and confirm the value b Adjust output to the reference value obtained using other means In tank level measurement if the actual level cannot be brought to zero for zero adjustment then the output ca...

Страница 45: ...er cover Screw each cover in tightly until it will not turn further There are two covers that must be locked on the ATEX Flameproof type transmitters An Allen head bolt shrouding bolt under the edge o...

Страница 46: ...as will be expelled in the direction of the arrow F0605 ai Figure 6 5 Draining Condensate from C10FR Flushing Connection Ring 6 5 2 Venting Gas from C10FR Flushing Connection Ring 1 Gradually open the...

Страница 47: ...CD update will be slower at low ambient temperature and it is recommended to use LPS function at temperatures above 10 C 6 6 1 Local Parameter Setting LPS Overview Parameter configuration by the zero...

Страница 48: ...the value Save Cancel the value 6 Output Mode 2 DP Unit 1 Tag Number 3 DP LRV 4 DP URV 7 Display Out1 8 LRV Apply Press 9 URV Apply Press 5 Damping Time Select Output mode Select DP unit Edit Tag numb...

Страница 49: ...below parameters are shown sequentially by each press of the push button Tag number Unit LRV URV Damping Output mode linear square root signal characterizer Display out 1 F0611 ai push push push push...

Страница 50: ...etween the various available pressure units F0612 ai screw push Select by the push button Blinking Blinking Save Cancel Available units kPa bar inH2O 4degC 39 2degF Torr mbar inH2O 20degC 68degF atm g...

Страница 51: ...or the differential prressure transmission part can be set Damping time constant is rounded off to two decimal places F0613 ai screw Change the first digit Set all other characters in the same way Set...

Страница 52: ...pletes the LRV setting Note 3 6 Press the push button The integral indicator then displays URV SET 7 Apply a pressure of 3 MPa to the transmitter Note 1 8 Turn the external adjustment screw in the des...

Страница 53: ...ter 6 6 13 Local Parameter Setting Lock To disable parameter changes by the Local Parameter Setting there are three different ways Locked features Communication Parameter Ext SW disable External Zero...

Страница 54: ...he slave transmitter for low pressure side measurement 7 1 Connection 7 1 1 Integral Indicator Display at Power On If the transmitter is equipped with the integral indicator when the EJXC40A digital r...

Страница 55: ...ansmitter and DD by following steps 1 and 2 If the appropriate DD is not installed in the configuration tool download it from the official HARTAssociation website or contact the configuration tool dis...

Страница 56: ...nsmission loop The configuration tool is connected in parallel with the transmitter but the polarity does not matter A connection example is shown below USB FieldMate Modem Relay terminals Distributor...

Страница 57: ...x y z SP represents one byte space 7 2 2 2 Setting Process Variables The transmitter can handle five types of data these are called device variables differential pressure DP module 1 side pressure Pr...

Страница 58: ...Measuring Range These range values are set as specified in the order before the instrument is shipped Follow the procedure below to change the measuring range You can enter the low limit LRV and high...

Страница 59: ...ime constant follow the procedure below Recalling the damping time constant parameter DP damp Recall parameters Root menu Detailed setup Sensors DP setup DP damp Enter a valid time constant in the DP...

Страница 60: ...the low cut value for the output signal is 1 and that for the integral indicator display is 10 which is equivalent to the square root of the low cut value Note that if the output mode is set to Sq ro...

Страница 61: ...pressure units are shown below inH2O 68degF inHg ftH2O 68degF mmH2O 68degF mmHg psi bar mbar g cm2 kg cm2 Pa kPa torr atm MPa inH2O mmH2O ftH2O hPa 2 Setting the low limits Pres 1 LRV Pres 2 LRV and h...

Страница 62: ...rameters Recall parameters Root menu Detailed setup Signal condition Pres 1 2 Swap Select Normal or Reverse 7 2 3 4 Bi directional Flow Measurement Setup This parameter enables output whose positive a...

Страница 63: ...ke sure that the upper limit and lower limit satisfy the following relationship Lower limit Upper limit 7 2 3 6 Integral Indicator Display Mode The mode setting for the output signal see subsection 7...

Страница 64: ...age PV Differential pressure DP Module 1 side pressure Pres 1 Module 1 side capsule temperature Temp 1 Module 2 side pressure Pres 2 Module 2 side capsule temperature Temp 2 User set scale PV value En...

Страница 65: ...following when entering the unit directly Up to eight characters can be set using alphanumeric characters and a slash but only the first six characters are displayed If two or more slashes or any of t...

Страница 66: ...suring range is zero Recalling and setting the zero point adjustment parameter DP Zero trim Recall parameters Root menu Maintenance DP trim DP Zero trim Confirm the value after it has stabilized ON th...

Страница 67: ...g and setting automatic adjustment parameters Recall parameters Root menu Maintenance DP trim DP trim method Auto Lower Pt Set the low limit value you want to adjust Auto Upper Pt Set the high limit v...

Страница 68: ...ance Pres 1 trim Clear Pres 1 trim method Pres 2 Root menu Maintenance Pres 2 trim Clear Pres 2 trim method 7 2 3 10 Analog Output Adjustment This function adjusts the analog output value Two adjustme...

Страница 69: ...outputs a 110 or higher signal For models with optional code C1 and C2 The switch is set to low L If an error occurs the transmitter outputs a 2 5 or lower signal The 4 to 20 mADC burnout direction f...

Страница 70: ...and Y coordinate values Enter the values to assign to the X and Y coordinates using the Point setting parameter 3 Apply the signal characterizer function After setting the coordinate values setting p...

Страница 71: ...tailed setup Output condition Process Alerts DPAlert Select differential pressure and set the threshold Pres 1 Alert Select module 1 side pressure and set the threshold Pres 2 Alert Select module 2 si...

Страница 72: ...d To perform test output from the DTM recall Loop Test from Hot key menu select manual test or auto test in the figure on the screen and set the value 2 Device variable simulation A value and status o...

Страница 73: ...ared F0711 ai Figure 7 2 9 Display during Squawk Execution 7 2 3 17 Burst Mode If burst mode is enabled the transmitter can continuously send three of the data shown in Table 7 2 4 through HART commun...

Страница 74: ...lect up to 8 from differential pressure pressure capsule temperature percent of range and Loop current 2 Cmd9 Device vars w status Continuous Window Process variable assigned to the head of Burst Vari...

Страница 75: ...d Burst Command Burst mode Burst Msg Trigger Mode Continuous Window Rising Falling On change Send to transmitter Cmd9 Cmd33 a Burst Command b Burst Variables c Update period and Max Update Period by S...

Страница 76: ...parameters Burst Condition Burst Message1 2 or 3 Burst Variables Burst Variable Code DP Select the differential pressure Pres 1 Select the module 1 side pressure Pres 2 Select the module 2 side pressu...

Страница 77: ...n Changes to device settings and changes to the device status through self diagnosis can be detected as events and alarm signals can be sent continuously Up to four events that occur are saved as hist...

Страница 78: ...tification Control to Enable event notification on token passing data link layer 3 2 Acknowledging events DTM only When the host device acknowledges an event event transmission stops To acknowledge ev...

Страница 79: ...nected to a single communication line To enable multidrop communication the device address must be set to any number ranging from 1 to 63 When set to multidrop mode all data are sent digitally Therefo...

Страница 80: ...earches for devices set to multidrop mode If the HART configuration tool is connected to devices their polling addresses and tags are displayed 2 Selecting the device of your choice enables communicat...

Страница 81: ...ess status Status group 5 14 Module 1 side setting status Status group 15 16 Diagnostic status Status group 17 Simulation status Status group 18 Module 2 side hardware status Status group 19 20 21 Mod...

Страница 82: ...ime Stamp Current Date The number of running days is displayed January 1 1970 is the operation start date Current Time The running time is displayed NOTE Timestamps are reset when the power is turned...

Страница 83: ...groups are displayed simultaneously 7 3 2 Advanced Diagnostics 7 3 2 1 What Is the Multi Sensing Process Monitoring Function The multi sensing process monitoring function optional code DG6 is a funct...

Страница 84: ...in the BlkF values or module 1 side fluctuation The blockage level obtained by statistically comparing the pressure fluctuation between the module 1 side and module 2 side For details see subsection 7...

Страница 85: ...o fP1 A value obtained by comparing the root mean square RMS of the module 1 side pressure fluctuation It is used for blockage detection Ratio fP1 SQRT fP1 Ref fP1 BlkF A value indicating the blockage...

Страница 86: ...ou are using this for the first time use the default values If not enough pressure fluctuation can be detected in valve simulation tests or in actual operation or if alarms occur frequently refer to s...

Страница 87: ...t Pres 2 low pressure side pressure fluctuation error is detected Determination of Pres 1 Side Blocking The BlkF value is prioritized in the determination of high pressure side blockage However if the...

Страница 88: ...on fP2 and Ref fP2 Pres 1 Side Blocking detection based on fP1 and Ref fP1 Pres 1 Pres 2 Side Blocking detection based on BlkF Ref fDP OK Ref fP2 OK Ref fP1 OK Ref BlkF NG Low pressure side blockage...

Страница 89: ...complete Diag Mode returns to the original mode NOTE If the impulse line blockage detection ILBD parameter is set using the Download to device of the DTM Diag Mode is automatically set to Stop Diag Pe...

Страница 90: ...t threshold e g Lim fDPmin Alarm occurrence Alarm occurrence Time Time Figure 7 3 4 Relationship between Diag Supp Count and alarms The detection count applies to all error alarms The factory default...

Страница 91: ...ILBD 1 2 Alarm masking Set the alarm to indicate through communication or on the LCD using the Diag Option parameter 3 1 Acquire reference values Acquire fluctuation data under steady state operation...

Страница 92: ...Pres 1 10 Invalid Ref Pres 2 11 Invalid Ref DP 12 ILBD over Range 13 Flg Temp Low Alarm Heat trace monitoring 14 Flg Temp High Alarm 15 Not used ILBD over range 1 Lim DPAvgmax The Lim DPAvgmax paramet...

Страница 93: ...ing 11 Invalid Ref DP To reflect alarms in analog alerts and LCD display follow the procedure below 1 Set the Diag Mode parameter to Stop 2 From the alarms assigned to bits 2 to 14 select the check bo...

Страница 94: ...s pressure fluctuation is small blockage detection does not work properly The pressure fluctuation may be small in the case of level measurement or pressure measurement of tanks or when gases are meas...

Страница 95: ...ference values automatically ends and the Diag Mode parameter switches automatically from Reference to Calculation Check that the sampling has finished by confirming that the Diag Mode has changed to...

Страница 96: ...ters elapses check that Pres 1 Side Blocking and Pres 2 Side Blocking are indicated in the Diag Error parameter 5 If the analog output is configured to shift in response to the alarm occurrence check...

Страница 97: ...ge the sampling period follow the procedure below 1 Set the Diag Mode parameter to Stop 2 Set the Diag Period parameter value in the range of 20 to 65535 seconds In addition increasing the detection c...

Страница 98: ...ion results Reference The reference values are set to the latest blockage detection data When the reference values are updated the mode automatically switches to Calculation 6 Diag Period 180 sec This...

Страница 99: ...ter can be entered only when Diag Mode is set to Stop 41 Lim BlkFmax Upper limit of one side blockage level BlkF used for impulse line blockage detection This parameter can be entered only when Diag M...

Страница 100: ...ected in Diag DP Comp NRatio fDP is used as the monitoring value NRatio fDP Sqrt fDP Ref fDP 53 NRatio fDP Status Status of NRatio fDP 54 Diag DPComp Compensation Flow rate compensation flag Sets whet...

Страница 101: ...High temperature error detection threshold Flg temp lo alert val Low temperature error detection threshold Formula Flg temp FT CT Cf CT AT CAUTION The flange temperature is calculated with the formula...

Страница 102: ...e case of DTM the tuning function can be used to set Flg temp coef Enter the measured flange temperature to calculate Flg temp coef using the capsule temperature and amplifier temperature measured wit...

Страница 103: ...put modes are Off Burnout and Fall back 4 Diag Fixed Out Val 21 6 mA A parameter for the fall back function in Diag Out option This specifies the output value of the 4 to 20mA analog signal when alarm...

Страница 104: ...rocedures required for component replacement Transmitters are precision instruments Please carefully and thoroughly read the following sections for information on how to properly handle them while per...

Страница 105: ...5 digits for 0 to 25 kPa 0 2 of rdg 0 1 of F S for 80 to 0 kPa 3 For 700 kPa class Accuracy 0 02 of rdg 3digits for 100 to 700 kPa 5 digits for 0 to 100 kPa 0 2 of rdg 0 1 of F S for 80 to 0 kPa 4 For...

Страница 106: ...e should be applied to the high pressure side as shown in the Figure The low pressure side is released to the atmosphere If the measurement range 0 point is 0 kPa or shifted in the positive direction...

Страница 107: ...lease pressures before disassembly Use proper tools for all operations Table 8 2 shows the tools required Table 8 2 Tools for Disassembly and Reassembly Tool Quantity Remarks Phillips screwdriver 1 JI...

Страница 108: ...re 8 3 4 Disconnect the DRS output cable cable with white connector at the end 5 Use a socket driver width across flats 5 5mm to loosen the two bosses 6 Carefully pull the CPU assembly straight forwar...

Страница 109: ...he Capsule Assembly IMPORTANT When cleaning the capsule assembly note the following items Handle the capsule assembly with care and be especially careful not to damage or distort the diaphragms that c...

Страница 110: ...transmitter Check correct operating conditions Measurement system problem F0805 ai YES NO NO YES Environmental conditions Operating conditions Transmitter 8 5 2 Troubleshooting Flowcharts YES NO NO Y...

Страница 111: ...DRS cable connected properly Check the installation of the impulse piping or process connections and take necessary measure to avoid leakage Are there any pressure leakage observed YES NO YES NO NO YE...

Страница 112: ...Amplifier temperature sensor failed AL 02 Amp 1 EEPROM Failure AL 02 Module 1 Amplifier EEPROM failed AL 02 CPU 1 Board Failure AL 02 Module 1 CPU board failed AL 02 1 AL 03 CAP2 ER F Pres 2 Sensor F...

Страница 113: ...he variable range For limit values see subsection 3 3 1 AL 12 TEMP1 S Temp 1 outside Limit AL 12 Module 1 Capsule temperature is outside range 50 to 130 degC AL 12 Continues to output the current valu...

Страница 114: ...Check the input value 4 AL 36 DP LO N DP Low Alarm AL 36 Differential pressure exceeds the specified lower limit threshold AL 36 AL 37 P1 HI N Pres 1 High Alarm AL 37 Module 1 Input pressure exceeds t...

Страница 115: ...J C Pres 1 Zero Trim Err AL 55 Module 1 Specified value is outside of setting range AL 55 Continues to output the current value Adjust so that it is within the specification range 14 Pres 1 Span Trim...

Страница 116: ...reach the reference level required to blockage detection AL 88 AL 88 INVR P2 C Invalid Ref Pres 2 AL 88 Module 2 Pressure fluctuation does not reach the reference level required to blockage detection...

Страница 117: ...mode for module 1 side pressure AL 91 AL 91 P2 SIM C Pres 2 Simulate Mode AL 91 Under simulation mode for module 2 side pressure AL 91 AL 91 T1 SIM C Temp 1 Simulate Mode AL 91 Under simulation mode...

Страница 118: ...current mode is desired but not set in that mode Set in the fixed current mode In multidrop mode Operation is set in the multi drop mode Not write protect mode Operation is set without a password Low...

Страница 119: ...corrupt Volatile Memory Defect 0x04 F Not used Watchdog Reset Executed 0x08 F Not used Power Supply Conditions Out of Range 0x10 S Not used Environmental Conditions Out of Range 0x20 S An internal or...

Страница 120: ...Good Not Limited Value Hold value Status Bad Low Limited High Limited 1 Temp 1 Sensor Failure AL 01 F Value Measured value Status Poor Accuracy Not Limited Value Measured value Status Bad Not Limited...

Страница 121: ...imited 1 Temp 2 Sensor Failure AL 03 F Value Measured value Status Poor Accuracy Not Limited Value Measured value Status Poor Accuracy Not Limited Value Measured value Status Bad Not Limited Cap 2 EEP...

Страница 122: ...t Limited AL 12 TEMP1 Temp 1 outside Limit AL 12 Non PV Out of Limits 0x02 S Value Measured value Status Poor Accuracy Not Limited AL 13 AMP TMP Amp T outside Limit AL 13 S Value Measured value Status...

Страница 123: ...L 35 DP HI DP High Alarm AL 35 N Value Measured value Status Good Not Limited AL 36 DP LO DP Low Alarm AL 36 N AL 37 P1 HI Pres 1 High Alarm AL 37 N AL 38 P1 LO Pres 1 Low Alarm AL 38 N AL 39 TMP1 HI...

Страница 124: ...of other than PV Value Measured value Status Good Not Limited DP Span Trim Err AL 53 C AL 54 P1 RNG Illegal Pres 1 LRV AL 54 C Value Measured value Status Good Not Limited In the case of PV Value Hold...

Страница 125: ...Limited In the case of PV Value Measured value Status Poor Accuracy Not Limited In the case of other than PV Value Measured value Status Good Not Limited Pres 2 Span Trim Err AL 57 C Pres 2 Fixed Mode...

Страница 126: ...valid Ref DP AL 88 Function Check 0x20 C Value Measured value Status Good Not Limited AL 88 INVR P1 Invalid Ref Pres 1 AL 88 C AL 88 INVR P2 Invalid Ref Pres 2 AL 88 C AL 88 INVR F Invalid Ref BlkF AL...

Страница 127: ...us Good Not Limited AL 91 P2 SIM Pres 2 Simulate Mode AL 91 C Value Measured value Status Good Not Limited Value and Status Simulated value Value Measured value Status Good Not Limited AL 91 T1 SIM Te...

Страница 128: ...tes the analog channel flag 0x00 R Bi direction mode Bi dir mode Bidirectional mode Sets bidirectional flow mode on or off On Off Off W Burst mode Burst mode Burst mode Switches to or out of burst mod...

Страница 129: ...update time for when there are no events and the minimum event retention time M Event Notification Retry Time Event Notification Retry Time Indicates the retry time for when events occur 60 000 s R Ma...

Страница 130: ...s diaphragm seal quantity information W Process Conn matl Process Connection material Stores or indicates process Connection material information W Process Conn size Process Connection size Stores or...

Страница 131: ...nge value of the user set PV scale As specified W Modify Engr unit User set PV scale value unit Edits the user set PV scale unit string M DP disp point Decimal point of the differential pressure displ...

Страница 132: ...m for module 1 side pressure Off Hi Al Detect Lo Al Detect Hi Lo Al Detect Off W Pres 1 Hi alert val High alert value for module 1 side pressure Sets the threshold for the high limit side alarm for mo...

Страница 133: ...capsule temperature R Temp 2 Limit Status Device variable limit status Indicates the limit status of the module 2 side capsule temperature R Temp 2 Data Quality Device variable data quality Indicates...

Страница 134: ...side zero adjustment point R Pres 1 UTD Module 1 side pressure upper trim deviation Indicates the module 1 side pressure s upper limit side span adjustment level R Pres 1 UTP Module 1 side pressure up...

Страница 135: ...e set R Pres 1 Min span Module 1 side minimum span Indicates the minimum span information of the module 1 side pressure R Pres 1 URV Upper range value of module 1 side pressure Sets the upper range va...

Страница 136: ...up 18 Hardware status information Indicates module 2 side hardware failure errors R Status group 19 Process status information Indicates module 2 side process errors outside the specification range er...

Страница 137: ...unit Sets the module 1 side pressure unit W Pres 2 unit Module 2 side pressure unit Sets the module 2 side pressure unit W Temp unit Temperature unit Sets the temperature unit W Write protection Enab...

Страница 138: ...ble DP Pres 1 Pres 2 Temp 1 Temp 2 M SV is SV assignment Indicates the variable currently assigned to SV Pres 1 R Change TV Assgn TV assignment Specifies the variable to assign to TV Teritary variable...

Страница 139: ...detecting Pres 2 Side Blocking using BlkF 0 60 W G Lim DPAvgmax DPAvg upper limit threshold Threshold for detection ILDB over Range using DPAvg and Invalid Ref DP using Ref DPAvg 1 00 W G Lim DPAvgmi...

Страница 140: ...Ref fP1 A value obtained by comparing the root mean square RSM of the module 1 side pressure fluctuation R G Ratio fP1 Status Status of Ratio fP1 R G BlkF A value obtained by comparing module 1 side p...

Страница 141: ...1EN EJX910A Multivariable Transmitter GS 01C25R01 01EN EJX930A Multivariable Transmitter GS 01C25R04 01EN EJXC50A EJXC40A EJAC50E C20FE C20FW C10FR EJXC80A EJAC80E C81FA C82FA C81FD C82FD C30SW C30SE...

Страница 142: ...parameters can be verified by reading these parameters from the optional local display or by checking actual calibration of the transmitter The calibration of the transmitter must be performed after...

Страница 143: ...ion manual should be followed The personnel performing the repair or replacement of the transmitter should have a sufficient skill level A1 2 7 Startup Time The transmitter generates a valid signal wi...

Страница 144: ...achinery plant apparatus under control of the system Basic Safety The equipment must be designed and manufactured such that it protects against risk of damage to persons by electrical shock and other...

Страница 145: ...ondition Check that the status of DP is GOOD Check the maximum and minimum values of DP Status Good DP Max Min Max 12 3 kPa Min 12 1 kPa 3 fDP under normal condition Check that the value of fDP is mor...

Страница 146: ...DP ILBD over range 8 ILBD parameters Record the values of parameters for ILBD operation Check the status of parameters for ILBD operation Record the value after checked that the status of each paramet...

Страница 147: ...ime Diag Period X Diag Supp Count passed Record the value after checked that the status is GOOD fDP 3 74856E 08 fP2 6 23277E 09 fP1 1 51409E 10 BlkF 0 8658873 Record the status of Checkbox in Diag Opt...

Страница 148: ...D over range Check that the alarm of Pres 2 Side Blocking is generated Check that the alarm of Pres 1 Side Blocking is not generated Diag Error Pres 2 Side Blocking Pres 1 Side Blocking 9 a 3 Both Sid...

Страница 149: ...value is closed Record the values of fDPand fP2 after the certain time Diag Period X Diag Supp Count passed Record the value after checked that the status is GOOD fDP 3 48562E 08 fP2 1 12328E 10 Recor...

Страница 150: ...d C30SW and C30SE 3rd Jan 2019 1 2 2 4 2 5 2 9 to 2 11 2 11 2 12 2 13 2 15 4 5 4 8 5 3 to 5 6 10 6 10 25 A1 1 to A1 3 Approved to IEC61508 by Notification body Add notes for keeping the manual and tag...