5-175
IM DLM4038-17EN
Commands
1
2
3
4
5
6
App
Index
:SERialbus<x>:CXPI:SEARch:SETup:ERRo
r:DLENgth
Function Sets or queries the CXPI bus signal search data
length error setting.
Syntax
:SERialbus<x>:CXPI:SEARch:SETup:
ERRor:DLENgth {<Boolean>}
:SERialbus<x>:CXPI:SEARch:SETup:
ERRor:DLENgth?
<x> = 1 to 4
Example
:SERIALBUS1:CXPI:SEARCH:SETUP:ERROR:
DLENGTH ON
:SERIALBUS1:CXPI:SEARCH:SETUP:ERROR:
DLENGTH?
-> :SERIALBUS1:CXPI:SEARCH:SETUP:
ERROR:DLENGTH 1
:SERialbus<x>:CXPI:SEARch:SETup:ERRo
r:FRAMing
Function Sets or queries the CXPI bus signal search
framing error setting.
Syntax
:SERialbus<x>:CXPI:SEARch:SETup:
ERRor:FRAMing {<Boolean>}
:SERialbus<x>:CXPI:SEARch:SETup:
ERRor:FRAMing?
<x> = 1 to 4
Example
:SERIALBUS1:CXPI:SEARCH:SETUP:ERROR:
FRAMING ON
:SERIALBUS1:CXPI:SEARCH:SETUP:ERROR:
FRAMING?
-> :SERIALBUS1:CXPI:SEARCH:SETUP:
ERROR:FRAMING 1
:SERialbus<x>:CXPI:SEARch:SETup:ERRo
r:IBS
Function Sets or queries the CXPI bus signal search IBS
error setting.
Syntax
:SERialbus<x>:CXPI:SEARch:SETup:
ERRor:IBS {<Boolean>}
:SERialbus<x>:CXPI:SEARch:SETup:
ERRor:IBS?
<x> = 1 to 4
Example
:SERIALBUS1:CXPI:SEARCH:SETUP:ERROR:
IBS ON
:SERIALBUS1:CXPI:SEARCH:SETUP:ERROR:
IBS?
-> :SERIALBUS1:CXPI:SEARCH:SETUP:
ERROR:IBS 1
:SERialbus<x>:CXPI:SEARch:SETup:ERRo
r:PARity
Function Sets or queries the CXPI bus signal search parity
error setting.
Syntax
:SERialbus<x>:CXPI:SEARch:SETup:
ERRor:PARity {<Boolean>}
:SERialbus<x>:CXPI:SEARch:SETup:
ERRor:PARity?
<x> = 1 to 4
Example
:SERIALBUS1:CXPI:SEARCH:SETUP:ERROR:
PARITY ON
:SERIALBUS1:CXPI:SEARCH:SETUP:ERROR:
PARITY?
-> :SERIALBUS1:CXPI:SEARCH:SETUP:
ERROR:PARITY 1
:SERialbus<x>:CXPI:SEARch:SETup:IDDa
ta?
Function Queries all ID and data condition settings for
CXPI bus signal searching.
Syntax
:SERialbus<x>:CXPI:SEARch:SETup:
IDData?
<x> = 1 to 4
:SERialbus<x>:CXPI:SEARch:SETup[:IDD
ata]:DATA?
Function Queries all data settings for the ID and data
conditions for CXPI bus signal searching.
Syntax
:SERialbus<x>:CXPI:SEARch:SETup[:
IDData]:DATA?
<x> = 1 to 4
:SERialbus<x>:CXPI:SEARch:SETup[:IDD
ata]:DATA:BCOunt
Function Sets or queries the pattern comparison position,
which is one of the ID and data conditions, for
CXPI bus signal searching.
Syntax
:SERialbus<x>:CXPI:SEARch:
SETup[:IDData]:DATA:BCOunt {<NRf>}
:SERialbus<x>:CXPI:SEARch:
SETup[:IDData]:DATA:BCOunt?
<x> = 1 to 4
<NRf> = 0 to 254
Example
:SERIALBUS1:CXPI:SEARCH:SETUP:
IDDATA:DATA:BCOUNT 1
:SERIALBUS1:CXPI:SEARCH:SETUP:
IDDATA:DATA:BCOUNT?
-> :SERIALBUS1:CXPI:SEARCH:SETUP:
IDDATA:DATA:BCOUNT 1
5.25 SERialbus Group