5-198
IM DLM4038-17EN
:SERialbus<x>:LIN:SEARch:SETup:IDDat
a:DATA:DECimal<y>
Function
Sets or queries the data value for LIN bus signal
searching in decimal notation.
Syntax
:SERialbus<x>:LIN:SEARch:SETup:
IDData:DATA:DECimal<y> {<NRf>}
:SERialbus<x>:LIN:SEARch:SETup:
IDData:DATA:DECimal<y>?
<x> = 1 to 4
<y> = 1 to 2
<NRf> = See the DLM4000 Features Guide for
this information.
Example
:SERIALBUS1:LIN:SEARCH:SETUP:IDDATA:
DATA:DECIMAL1 1
:SERIALBUS1:LIN:SEARCH:SETUP:IDDATA:
DATA:DECIMAL1?
-> :SERIALBUS1:LIN:SEARCH:SETUP:IDDA
TA:DATA:DECIMAL1 1.000E+00
:SERialbus<x>:LIN:SEARch:SETup:IDDat
a:DATA:ENDian
Function
Sets or queries the data endian setting for LIN
bus signal searching.
Syntax
:SERialbus<x>:LIN:SEARch:SETup:
IDData:DATA:ENDian {BIG|LITTle}
:SERialbus<x>:LIN:SEARch:SETup:
IDData:DATA:ENDian?
<x> = 1 to 4
Example
:SERIALBUS1:LIN:SEARCH:SETUP:IDDATA:
DATA:ENDIAN BIG
:SERIALBUS1:LIN:SEARCH:SETUP:IDDATA:
DATA:ENDIAN?
-> :SERIALBUS1:LIN:SEARCH:SETUP:IDDA
TA:DATA:ENDIAN BIG
:SERialbus<x>:LIN:SEARch:SETup:IDDat
a:DATA:HEXa<y>
Function
Sets the data for LIN bus signal searching in
hexadecimal notation.
Syntax
:SERialbus<x>:LIN:SEARch:SETup:
IDData:DATA:HEXa<y> {<String>}
<x> = 1 to 4
<y> = 1 to 8
Example
:SERIALBUS1:LIN:SEARCH:SETUP:IDDATA:
DATA:HEXA1 "12"
:SERialbus<x>:LIN:SEARch:SETup:IDDat
a:DATA:MODE
Function
Sets or queries the data enable/disable condition
for LIN bus signal searching.
Syntax
:SERialbus<x>:LIN:SEARch:SETup:
IDData:DATA:MODE {<Boolean>}
:SERialbus<x>:LIN:SEARch:SETup:
IDData:DATA:MODE?
<x> = 1 to 4
Example
:SERIALBUS1:LIN:SEARCH:SETUP:IDDATA:
DATA:MODE ON
:SERIALBUS1:LIN:SEARCH:SETUP:IDDATA:
DATA:MODE?
-> :SERIALBUS1:LIN:SEARCH:SETUP:IDDA
TA:DATA:MODE 1
:SERialbus<x>:LIN:SEARch:SETup:IDDat
a:DATA:MSBLsb
Function
Sets or queries the data MSB and LSB bits for
LIN bus signal searching.
Syntax
:SERialbus<x>:LIN:SEARch:SETup:
IDData:DATA:MSBLsb {<NRf>,<NRf>}
:SERialbus<x>:LIN:SEARch:SETup:
IDData:DATA:MSBLsb?
<x> = 1 to 4
<NRf> = 0 to 63
Example
:SERIALBUS1:LIN:SEARCH:SETUP:IDDATA:
DATA:MSBLSB 1,0
:SERIALBUS1:LIN:SEARCH:SETUP:IDDATA:
DATA:MSBLSB?
-> :SERIALBUS1:LIN:SEARCH:SETUP:IDDA
TA:DATA:MSBLSB 1,0
:SERialbus<x>:LIN:SEARch:SETup:IDDat
a:DATA:PATTern<y>
Function
Sets or queries the data value for LIN bus signal
searching in binary notation.
Syntax
:SERialbus<x>:LIN:SEARch:SETup:
IDData:DATA:PATTern<y> {<String>}
:SERialbus<x>:LIN:SEARch:SETup:
IDData:DATA:PATTern<y>?
<x> = 1 to 4
Example
:SERIALBUS1:LIN:SEARCH:SETUP:IDDATA:
DATA:PATTERN1 "00110101"
:SERIALBUS1:LIN:SEARCH:SETUP:IDDATA:
DATA:PATTERN1?
-> :SERIALBUS1:LIN:SEARCH:SETUP:IDDA
TA:DATA:PATTERN1 "00110101"
5.25 SERialbus Group