5-71
IM DLM4038-17EN
Commands
1
2
3
4
5
6
App
Index
:ANALysis:PANalyze<x>:SWLoss:DPROxim
al:PERCent
Function
Sets or queries the distal, mesial, and proximal
points as percentages.
Syntax
:ANALysis:PANalyze<x>:SWLoss:DPROxim
al:PERCent {<NRf>,<NRf>,<NRf>}
:ANALysis:PANalyze<x>:SWLoss:DPROxim
al:PERCent?
<NRf>, <NRf>, <NRf> = 0 to 100 (%, in steps of 1)
Example
:ANALYSIS:PANALYZE1:SWLOSS:
DPROXIMAL:PERCENT 10,50,90
:ANALYSIS:PANALYZE1:SWLOSS:
DPROXIMAL:PERCENT?
-> :ANALYSIS:PANALYZE1:SWLOSS:DPROXI
MAL:PERCENT 10,50,90
:ANALysis:PANalyze<x>:SWLoss:DPROxim
al:UNIT
Function
Sets or queries the distal, mesial, and proximal
points as voltages.
Syntax
:ANALysis:PANalyze<x>:SWLoss:DPROxim
al:UNIT {<NRf>,<NRf>,<NRf>}
:ANALysis:PANalyze<x>:SWLoss:DPROxim
al:UNIT?
<NRf>, <NRf>, <NRf> = See the DLM4000
Features Guide for this information.
Example
:ANALYSIS:PANALYZE1:SWLOSS:
DPROXIMAL:UNIT −1,0,1
:ANALYSIS:PANALYZE1:SWLOSS:D
PROXIMAL:UNIT?
-> :ANALYSIS:PANALYZE1:SWLOSS:DPROXI
MAL:UNIT −1.000,
0.000,1.000
:ANALysis:PANalyze<x>:SWLoss:DTYPe
Function
Sets or queries the device type for total loss
computation.
Syntax
:ANALysis:PANalyze<x>:SWLoss:DTY
Pe {IGBT|MOSFET|OFF}
:ANALysis:PANalyze<x>:SWLoss:DTYPe?
<x> = 1 to 2
Example
:ANALYSIS:PANALYZE1:SWLOSS:DTY
PE IGBT
:ANALYSIS:PANALYZE1:SWLOSS:DTYPE?
-> :ANALYSIS:PANALYZE1:SWLOSS:DTY
PE IGBT
:ANALysis:PANalyze<x>:SWLoss:ILEVel
Function
Sets or queries the current level used to
determine the zero loss period for total loss
computation.
Syntax
:ANALysis:PANalyze<x>:SWLoss:ILEV
el {<NRf>|<Current>}
:ANALysis:PANalyze<x>:SWLoss:ILEVel?
<x> = 1 to 2
<NRf>, <Current> = See the DLM4000 Features
Guide for this information.
Example
:ANALYSIS:PANALYZE1:SWLOSS:ILEVEL 1
:ANALYSIS:PANALYZE1:SWLOSS:ILEVEL?
-> :ANALYSIS:PANALYZE1:SWLOSS:ILEVE
L 1.000
:ANALysis:PANalyze<x>:SWLoss:MATH
Function
Sets or queries the on/off status of the power
waveform display.
Syntax
:ANALysis:PANalyze<x>:SWLoss:MATH {O
FF|POWer}
:ANALysis:PANalyze<x>:SWLoss:MATH?
<x> = 1 to 2
Example
:ANALYSIS:PANALYZE1:SWLOSS:MATH OFF
:ANALYSIS:PANALYZE1:SWLOSS:MATH?
-> :ANALYSIS:PANALYZE1:SWLOSS:MATH O
FF
:ANALysis:PANalyze<x>:SWLoss:MEASu
re?
Function
Queries all of the settings for the automated
measurement of power supply analysis
parameters.
Syntax
:ANALysis:PANalyze<x>:SWLoss:MEASu
re?
<x> = 1 to 2
:ANALysis:PANalyze<x>:SWLoss:MEASure
:<Parameter>?
Function
Queries the setting of a power supply analysis
parameter.
Syntax
:ANALysis:PANalyze<x>:SWLoss:MEASure
:<Parameter>?
<x> = 1 to 2
<Parameter> =
{P|PABS|PN|PP|WH|WHABs|
WHN|WHP|Z}
:ANALysis:PANalyze<x>:SWLoss:MEASure
:<Parameter>:COUNt?
Function
Queries the continuous statistical processing
count of a power supply analysis parameter.
Syntax
:ANALysis:PANalyze<x>:SWLoss:
MEASure:<Parameter>:COUNt?
<x> = 1 to 2
<Parameter> =
{P|PABS|PN|PP|WH|WHABs|
WHN|WHP|Z}
Example
:ANALYSIS:PANALYZE1:SWLOSS:MEASURE:
P:COUNT?
-> :ANALYSIS:PANALYZE1:SWLOSS:MEASUR
E:P:COUNT 100
5.3 ANALysis Group