5-130
IM DLM4038-17EN
:LOGic:PODL:BIT<x>?
Function
Queries all settings of a bit of logic input port L.
Syntax
:LOGic:PODL:BIT<x>?
<x> = 0 to 7
:LOGic:PODL:BIT<x>:DISPlay
Function
Sets or queries the on/off status of all bit displays
of logic input port L.
Syntax
:LOGic:PODL:BIT<x>:DISPlay {<Boolean
>}
:LOGic:PODL:BIT<x>:DISPlay?
<x> = 0 to 7
Example
:LOGIC:PODL:BIT1:DISPLAY ON
:LOGIC:PODL:BIT1:DISPLAY?
-> :LOGIC:PODL:BIT1:DISPLAY 1
:LOGic:PODL:BIT<x>:LABel
Function
Sets or queries the label of a bit of logic input port
L.
Syntax
:LOGic:PODL:BIT<x>:LABel {<String>}
:LOGic:PODL:BIT<x>:LABel?
<x> = 0 to 7
<String> = Up to eight characters
Example
:LOGIC:PODL:BIT1:LABEL "ABC"
:LOGIC:PODL:BIT1:LABEL?
-> :LOGIC:PODL:BIT1:LABEL "ABC"
:LOGic:PODL:BIT<x>:LEVel
Function
Sets or queries the user-defined threshold level
of a bit of logic input port L.
Syntax
:LOGic:PODL:BIT<x>:LEVel {<Voltage>}
:LOGic:PODL:BIT<x>:LEVel?
<x> = 0 to 7
<Voltage> = −10 to 10 V (in 0.1-V steps)
Example
:LOGIC:PODL:BIT1:LEVEL 1
:LOGIC:PODL:BIT1:LEVEL?
-> :LOGIC:PODL:BIT1:LEVEL 1.000E+00
:LOGic:PODL:BIT<x>:TYPE
Function
Selects the threshold level of a bit of logic input
port L.
Syntax
:LOGic:PODL:BIT<x>:TYPE {CMOS1|CMOS2
|CMOS3|CMOS5|ECL|USERdefine}
:LOGic:PODL:BIT<x>:TYPE?
<x> = 0 to 7
Example
:LOGIC:PODL:BIT1:TYPE CMOS1
:LOGIC:PODL:BIT1:TYPE?
-> :LOGIC:PODL:BIT1:TYPE CMOS1
:LOGic:PODL:BITOrder
Function
Sets or queries the bit order of logic input port L.
Syntax
:LOGic:PODL:BITOrder {<String>}
:LOGic:PODL:BITOrder?
<String> = Combination of “L0” to “L7” and “BUS”
(19 characters)
Example
:LOGIC:PODL:BITORDER "L0L1L2L3L4L5L6
L7BUS"
:LOGIC:PODL:BITORDER?
-> :LOGIC:PODL:BITORDER "L0L1L2L3L4L
5L6L7BUS"
:LOGic:PODL:BUS?
Function
Queries all bus settings of a bit of logic input port L.
Syntax
:LOGic:PODL:BUS?
:LOGic:PODL:BUS:ASSignment
Function
Sets or queries the assignment of a bus of logic
input port L.
Syntax
:LOGic:PODL:BUS:
ASSignment {<String>}
:LOGic:PODL:BUS:ASSignment?
<String> = Up to 16 characters
Example
:LOGIC:PODL:BUS:ASSIGNMENT "L1"
:LOGIC:PODL:BUS:ASSIGNMENT?
-> :LOGIC:PODL:BUS:ASSIGNMENT "L1"
Description <String> = Combination of “L0” to “L7”
:LOGic:PODL:BUS:DISPlay
Function
Sets or queries the bus display on/off status of
logic input port L.
Syntax
:LOGic:PODL:BUS:DISPlay {<Boolean>}
:LOGic:PODL:BUS:DISPlay?
Example
:LOGIC:PODL:BUS:DISPLAY ON
:LOGIC:PODL:BUS:DISPLAY?
-> :LOGIC:PODL:BUS:DISPLAY 1
:LOGic:PODL:BUS:FORMat
Function
Sets or queries the bus display format of logic
input port L.
Syntax
:LOGic:PODL:BUS:FORMat {HEX|BINary}
:LOGic:PODL:BUS:FORMat?
Example
:LOGIC:PODL:BUS:FORMAT HEX
:LOGIC:PODL:BUS:FORMAT?
-> :LOGIC:PODL:BUS:FORMAT HEX
:LOGic:PODL:BUS:LABel
Function
Sets or queries the label of a bus of logic input
port L.
Syntax
:LOGic:PODL:BUS:LABel {<String>}
:LOGic:PODL:BUS:LABel?
<String> = Up to eight characters
Example
:LOGIC:PODL:BUS:LABEL "PODL"
:LOGIC:PODL:BUS:LABEL?
-> :LOGIC:PODL:BUS:LABEL "PODL"
5.19 LOGic Group