5-143
IM DLM4038-17EN
Commands
1
2
3
4
5
6
App
Index
:MEASure:CYCLe:TRACe
Function
Sets or queries the source waveform used in
the cyclic statistical processing for automated
measurement.
Syntax
:MEASure:CYCLe:TRACe {OWN|<NRf>|
MATH<x>|PODA<x>|PODB<x>|PODL<x>}
:MEASure:CYCLe:TRACe?
<NRf> = 1 to 8
<x> of MATH<x> = 1 to 4
<x> of PODA<x>, PODB<x>, PODL<x> = 0 to 7
Example
:MEASURE:CYCLE:TRACE 1
:MEASURE:CYCLE:TRACE?
-> :MEASURE:CYCLE:TRACE 1
:MEASure:HISTory:ABORt
Function
Aborts statistical processing of automatically
measured history waveforms.
Syntax
:MEASure:HISTory:ABORt
Example
:MEASURE:HISTORY:ABORT
:MEASure:HISTory:EXECute
Function
Executes statistical processing of automatically
measured history waveforms.
Example
:MEASURE:HISTORY:EXECUTE
:MEASure:INDicator
Function
Sets or queries the measurement location
indicator.
Syntax
:MEASure:INDicator {<NRf>|MATH<x>|
PODA<x>|PODB<x>|PODL<x>|
OFF{,<Parameter>[,2]}}
:MEASure:INDicator?
<NRf> = 1 to 8
<x> of MATH<x> = 1 to 4
<x> of PODA<x>, PODB<x>, PODL<x> = 0 to 7
<Parameter> =
{AMPLitude|AVERage|
AVGFreq|AVGPeriod|BWIDth|DELay|
DUTYcycle|FALL|FREQuency|HIGH|LOW|
MAXimum|MINimum|NOVershoot|NWIDth|
P|PABS|PERiod|PN|POVershoot|PP|
PTOPeak|PWIDth|RISE|RMS|V1|V2|WH|
WHABs|WHN|WHP|Z}
Example
:MEASURE:INDICATOR 1,AVERAGE
:MEASURE:INDICATOR?
-> :MEASURE:INDICATOR 1,AVERAGE
:MEASure:MODE
Function
Sets or queries the on/off/statistical processing
status of automated measurement.
Syntax
:MEASure:MODE {OFF|ON|CONTinuous|
CYCLe|HISTory}
:MEASure:MODE?
Example
:MEASURE:MODE OFF
:MEASURE:MODE?
-> :MEASURE:MODE OFF
:MEASure:{PODA<x>|PODB<x>|PODL<x>}?
Function
Queries all the parameter on/off status of the
logic waveform.
Syntax
:MEASure:{PODA<x>|PODB<x>|PODL<x>}?
<x> = 0 to 7
:MEASure:{PODA<x>|PODB<x>|PODL<x>}
{[:AREA1]|:AREA2}?
Function
Queries all the parameter on/off status of the
logic waveform in the specified area.
Syntax
:MEASure:{PODA<x>|PODB<x>|PODL<x>}
{[:AREA1]|:AREA2}?
<x> = 0 to 7
:MEASure:{PODA<x>|PODB<x>|PODL<x>}
{[:AREA1]|:AREA2}:ALL
Function
Collectively turns on or off all measurement items
of a logic waveform in the specified area.
Syntax
:MEASure:{PODA<x>|PODB<x>|PODL<x>}
{[:AREA1]|:AREA2}ALL {<Boolean>}
<x> = 0 to 7
Example
:MEASURE:PODL1:AREA1:ALL ON
:MEASure:{PODA<x>|PODB<x>|PODL<x>}
{[:AREA1]|:AREA2}:<Parameter>?
Function
Queries the setting of a logic waveform parameter
in the specified area.
Syntax
:MEASure:{PODA<x>|PODB<x>|PODL<x>}
{[:AREA1]|:AREA2}:<Parameter>?
<x> = 0 to 7
<Parameter> =
{AVGFreq|DELay|DUTYcycle
|FREQuency|PERiod|PNUMber}
:MEASure:{PODA<x>|PODB<x>|PODL<x>}{[
:AREA1]|:AREA2}:<Parameter>:COUNt?
Function
Queries the statistical processing count of a logic
waveform parameter in the specified area.
Syntax
:MEASure:{PODA<x>|PODB<x>|PODL<x>}{[
:AREA1]|:AREA2}:<Parameter>:COUNt?
<x> = 0 to 7
<Parameter> =
{AVGFreq|DELay|DUTYcycle
|FREQuency|PERiod|PNUMber}
Example
:MEASURE:PODL:AREA1:AVGFREQ:COUNT?
-> :MEASURE:PODL:AREA1:AVGFREQ:COUN
T 100
5.21 MEASure Group