4-12
IM 701310-01E
Setup Data after Executing Auto Setup
Related CH1 to CH4
Select
INPUT
Position
0 div
Coupling
DC1 MΩ except DC50 Ω
No change for DC50 Ω
BW
FULL
Offset
0 V
Invert
OFF
M1 to M4
Display
OFF
Acquisition
Mode
Normal
Hireso
OFF
SAMPLING/LENGTH
Interp
ON
Repetitive
OFF
Interleave
OFF
Length
125 kW
Trigger
Mode
Auto
HoldOff
Min (20 ns)
Delay
OFF
Position
50%
Type
Edge
Polarity
Rise
Coupling
DC
Hysteresis
Small
HF Rej
OFF
Window
OFF
Accumulation
Mode
OFF
Screen display
Mapping
Auto
Dot Connect
Sine
Intensity
10 (Default)
Brightness
8 (Default)
Zoom
Zoom
OFF
Main
ON
Cursor
Display
OFF
Waveform parameter
Display
OFF
Telecom Test
Display
OFF
Analysis
Display
OFF
Items depending on input signals
CH On/Off
ON when detecting voltage of �7 mV or higher, OFF when detecting
less than �7 mV
V/div
Select the highest range of sensitivity that does not exceed �3.5 div
Trigger Level
Center
Trigger Source
Channel whose input amplitude (Max-Min) is 1 division or greater with
the lowest frequency
T/Div
The fastest sweep range that exceeds 5 ms/div whose input
amplitude is 1 division or greater and which the fastest frequency can
be observed with 2 cycles or greater
FORM VT Form
1 to 4 windows based on the active channels
4.5 Performing Auto Setup