6-26
IM 701310-01E
Trigger Level
This is the same as the edge trigger. See section 6.6 for details.
Example
Trigger source
More than
Trigger
t
T1
t > T1
Trigger source
Less than
Trigger
t
T1
t < T1
Trigger source
Between
Trigger
t
T1
T1 < t < T2
T2
Trigger source
Trigger
t
T1
t > T1
Time out
Trigger source
Out of Range
Trigger
t
T1
T2
Trigger
t < T1 or t > T2
The following figure shows how the trigger point differs between More than and Time
Out.
More than
Time = 400 ns
Trigger
500 ns
100 ns
CH1
Trigger
500 ns
100 ns
CH1
400 ns
Time Out
Time = 400 ns
6.10 Activating a Trigger on a Pulse Width