10-66
IM 701310-01E
When the Width Type Is State
• Setting the Clock Source, State Condition, and Logic
8.
Press the
State
soft key.
9.
Press the
Clock
soft key.
10.
Press the soft key that corresponds to the appropriate clock source waveform.
To select a channel from M1 to M4, press the
M1-M4
soft key first.
11.
Press the appropriate waveform soft key to select the status.
• To select a channel from M1 to M4, press the
M1-M4
soft key first.
• If the waveform is selected for the clock source in step 9, select the slope.
• H: High. L: Low. X: Don’t care
12.
Press the
Logic
soft key to select AND or OR.
13.
Press
ESC
to return to the previous screen.
Example in which
the clock source
is set to CH1
• Selecting the False-to-True Condition Change or the True-to-False Condition
Change
14.
Press the
Polarity
soft key to select True or False.
• Setting the Level and Hysteresis
15.
Press the
Level/Hys
soft key.
16.
Press the appropriate waveform soft key, and then use the
rotary knob
to set the
status reference level and hysteresis.
• To select a channel from M1 to M4, press the
M1-M4
soft key first.
• Press the soft key to select the setting that you want to set using the rotary knob.
17.
Press
ESC
to return to the previous screen.
Level
Hys
• Executing the Search
Carry out steps 6 to 14 in section 10.10.
10.11 Searching Waveform Signals