10-60
IM 701310-01E
Executing the Search
12.
Press the
Exec
soft key. A section that meets the search conditions appears in
the zoom window.
The Exec soft key changes to the Abort soft key. To stop the search, press the
Abort
key.
Displaying Search Results
13.
Use the
rotary knob
to select a found-point number.
The waveform that corresponds to the number appears in the zoom waveform area.
Turning a Found-Point Mark ON and OFF
14.
Press the
Pattern#/Mark
soft key to select ON or OFF.
Found-point number
Turn found-point marks ON and OFF
You can display found-point marks at the top edge of the main and zoom windows.
Found-point marks indicate the waveform positions that were found. The found-
point mark that matches the found-point number is highlighted.
Explanation
The DL9000 searches the displayed waveform for sections that meet certain conditions
and displays the found results in the zoom window.
Search Type
• The DL9000 can search a serial bus signal (Serial, I2C, CAN, LIN, SPI, or UART).
• You can set edge, edge (qualified), state, and width conditions to search an waveform
signal.
Search Start Point
The selectable range is �5.00 divisions. The resolution is 0.01 divisions.
Skip Mode
After finding a point that meets the search conditions, the DL9000 skips searching for
the specified time or count.
OFF
Searches all found points.
Hold Off
Skips searching for the specified time.
The selectable range is 0.1 ns to 1.00000 s (six significant digits). The resolution
is 0.1 ns.
Decimation
Skips searching for the specified count.
The selectable range is 1 to 9999.
Displaying Search Results
Numbers are assigned to the points that are found. Zero is assigned to the first found
point, one is assigned to the second found point, and so on.
• The maximum found-point number is 4999.
• You can display the waveform that corresponds to the selected found-point number in
the zoom waveform area.
10.10 Selecting the Search Type and Skip Mode, Executing the Search, and Displaying the Results