10-20
IM 701310-01E
Setting the Threshold Levels for Measurement Items Other Than Rise and
Fall
36.
Press the
Threshold Setup
soft key.
Selecting the Source Waveform
37.
Press the source waveform soft key.
Selecting the Method for Setting the Threshold Level
38.
Press the
Threshold
soft key.
39.
Press the Auto, Level/Hys, or Upper/Lower soft key to select method.
• If you select Auto, proceed to step 40.
• If you select Level/Hys, proceed to step 41.
• If you select Upper/Lower, proceed to step 43.
• Auto
40.
Press the
Center of
soft key to select P-P or High-Low.
• Level/Hys
41.
Press the
Level/Hys
soft key.
42.
Use the
rotary knob
to set the threshold level and hysteresis.
Press the soft key to select the setting that you want to set using the rotary knob.
• Upper/Lower
43.
Press the
Upper/Lower
soft key.
44.
Use the
rotary knob
to set the upper and lower limits.
Press the soft key to select the limit that you want to set using the rotary knob.
Auto
Level/Hys
Upper/Lower
Level
Hys
Upper
Lower
10.2 Performing Automated Measurement of Waveform Parameters