<4. Measurement Procedures>
4-17
IM 80J01A01-01E
4.1.3
Procedures of Setting Gas Mixer
This procedure is only perfomed when controlling environment around
sample. If you
don’t need environment control, go to procedure shown
in
Start Gas Mixer by referring to
3.3 Starting and Shutting Down of Gas
In this section, procedures are described for example by setting as
following.
- Sample vessel: microplate
- CO
2
concentration in chamber: 5 %
- O
2
concentration in chamber: atomostic concentration (
≓
20 %)
For other settings, please refer to Technical Information TI
80J01A05-01E (Setting Parameter of Gas Mixer).
By
“Up/ Down” key of “GAS FLOW”, set mixture gas flow rate to 150
(ml / min) and push
“Enter”.
By “Up/ Down” key of “CO
2
CONCENTRATION
”, set CO
2
concentration
to 6.0 (%).
By “Up/ Down” key of “O
2
CONCENTRATION
”, set O
2
concentration to 20.0 (%).
Содержание CQ1
Страница 2: ...This document corresponds to the following versions of software Measurement software R1 04 ...
Страница 8: ...vi IM 80J01A01 01E Right Side of Utility Box Aperture label ...
Страница 24: ......
Страница 34: ......
Страница 48: ......
Страница 177: ... 5 Functions of CQ1 Software 5 81 IM 80J01A01 01E Drag tab to move the window ...
Страница 178: ...5 82 5 Functions of CQ1 Software IM 80J01A01 01E Drag tab to square mark as following to add the tab to the window ...
Страница 187: ... 5 Functions of CQ1 Software 5 91 IM 80J01A01 01E 4 Area Window Display image of Area by tile Change to Area from Field ...
Страница 236: ......
Страница 238: ......
Страница 252: ...7 14 7 Trouble Shooting IM 80J01A01 01E Set sub cover and tighten 2 screws ...
Страница 256: ......
Страница 266: ......