<5. Functions of CQ1 Software>
5-55
IM 80J01A01-01E
(3) XY Stage
Set moving speed of XY stage. [Edit] button is to select XY stage
speed.
①
Speed
Select XY stage speed. Default setting is the fastest. If samples
are easy to ablate from sample vessel, slow setting is
recommended.
To apply, click
③
Set
②
Cancel
Cancel XY stage speed change
③
Set
Set XY stage speed
(3) Laser-Power Calibration
Click
“Calibrate” button to show “Laser-Power Calibration” dialog, and
click
“Yes” to proceed calibration, but no calibration by clicking “No”.
Laser power calibration is performed to maintain excitation laser power
irradiated to samples against
“Excitation Power” set in “Imaging”
window (refer to
). It takes a few minutes per 1
laser line to calibrate.
In case that error occurs in laser power calibration, optical adjustment
is needed. Please contact the dealer.
①
②
③
Содержание CQ1
Страница 2: ...This document corresponds to the following versions of software Measurement software R1 04 ...
Страница 8: ...vi IM 80J01A01 01E Right Side of Utility Box Aperture label ...
Страница 24: ......
Страница 34: ......
Страница 48: ......
Страница 177: ... 5 Functions of CQ1 Software 5 81 IM 80J01A01 01E Drag tab to move the window ...
Страница 178: ...5 82 5 Functions of CQ1 Software IM 80J01A01 01E Drag tab to square mark as following to add the tab to the window ...
Страница 187: ... 5 Functions of CQ1 Software 5 91 IM 80J01A01 01E 4 Area Window Display image of Area by tile Change to Area from Field ...
Страница 236: ......
Страница 238: ......
Страница 252: ...7 14 7 Trouble Shooting IM 80J01A01 01E Set sub cover and tighten 2 screws ...
Страница 256: ......
Страница 266: ......