<1. Function Overview>
1-3
IM 80J01A01-01E
1.2
Principle of the Nipkow Disk System with Micro-lens
Two disks, including the “pinhole array disk” having many pinholes
arranged
in a helical pattern, and the “Micro-lens array disk” that
condenses excitation laser to individual pinholes, are rotated jointly at
high speed to perform multiple scans over the observation area with
approximately 1,000 laser beams.
Multiple beam scans are performed not only at high speed, but also
with each beam exciting fluorochromes at high efficiency and very low
laser intensity. These results in an optimal live cell observation system
where phototoxicity and fluorescence photobleaching are suppressed
notably compared to any conventional systems while allowing
high-resolution real confocal imaging.
Operating Principle of the Micro-lens enhanced dual Nipkow Disk
System
Содержание CQ1
Страница 2: ...This document corresponds to the following versions of software Measurement software R1 04 ...
Страница 8: ...vi IM 80J01A01 01E Right Side of Utility Box Aperture label ...
Страница 24: ......
Страница 34: ......
Страница 48: ......
Страница 177: ... 5 Functions of CQ1 Software 5 81 IM 80J01A01 01E Drag tab to move the window ...
Страница 178: ...5 82 5 Functions of CQ1 Software IM 80J01A01 01E Drag tab to square mark as following to add the tab to the window ...
Страница 187: ... 5 Functions of CQ1 Software 5 91 IM 80J01A01 01E 4 Area Window Display image of Area by tile Change to Area from Field ...
Страница 236: ......
Страница 238: ......
Страница 252: ...7 14 7 Trouble Shooting IM 80J01A01 01E Set sub cover and tighten 2 screws ...
Страница 256: ......
Страница 266: ......