<7. Trouble Shooting>
7-7
IM 80J01A01-01E
Problem
Probable Cause / Corrective Action
Z position of bottom of sample vessel
can not be found.
⇒
Reset the range of
“Search”
function wider by using
“▼” button
beside
“Search” button.
⇒
Confirm that sample vessel is set
properly.
The sample can not be found.
It might be that there are not any cells
in F.O.V.
⇒
First, run
“Preview map image
acquisition
” function in “Map”
window, and, after that, reset the
Field position to region containing
cells according to the result of the
function described above.
Well(s) can not be selected by
“Imaging” / “Analysis” / “Well Table”
windows.
It might be that
“Select measurement
area” mode is ineffective.
⇒
Click
button to set the mode
effective.
Measurement and analysis can not
be run simultaneously.
It might be that
“CQ” mode is not
selected.
⇒
Select
“CQ” mode.
It might be that the check box of
“Execute Analysis” in “Application”
window is not checked.
⇒
Check the check box described
above.
Lasers do not operate.
It might be that the key switch switch
is not turned ON.
⇒
Turn it ON.
It might be that the remote interlock
connector is not connected to the
paired port.
⇒
Connect the connector to the
paired port.
It might be that the door is open.
⇒
Close it.
It might be that sub cover is open.
⇒
Close it.
⇒
Confirm ONLINE and Laser: ON in
status panel (refer to
There is(are) gap(s) in resultant
image of
“Map” window.
⇒
There is occasionally happened
that the gap appears according to
the magnification of resultant
image of
“Map” window.
There are dark fields sometimes in
map imaging.
⇒
Содержание CQ1
Страница 2: ...This document corresponds to the following versions of software Measurement software R1 04 ...
Страница 8: ...vi IM 80J01A01 01E Right Side of Utility Box Aperture label ...
Страница 24: ......
Страница 34: ......
Страница 48: ......
Страница 177: ... 5 Functions of CQ1 Software 5 81 IM 80J01A01 01E Drag tab to move the window ...
Страница 178: ...5 82 5 Functions of CQ1 Software IM 80J01A01 01E Drag tab to square mark as following to add the tab to the window ...
Страница 187: ... 5 Functions of CQ1 Software 5 91 IM 80J01A01 01E 4 Area Window Display image of Area by tile Change to Area from Field ...
Страница 236: ......
Страница 238: ......
Страница 252: ...7 14 7 Trouble Shooting IM 80J01A01 01E Set sub cover and tighten 2 screws ...
Страница 256: ......
Страница 266: ......