<7. Trouble Shooting>
7-15
IM 80J01A01-01E
7.3
Trouble Shooting during Analyzing
Problem
Probable Cause / Corrective Action
“Pre Analysis” function does not run
after the measurement.
“Pre Analysis” function does not run
in
“CQ”mode.
⇒
Change the oprerating mode from
“CQ” mode to “HDD” mode, and
then run
“Pre Analysis”.
Nothing is shown in Thumbnail Chart
window.
⇒
Select well to show chart in Well
Table window
Heat map display in Well Table
window disappeares.
Heat map
isn’t shown if wells are
selected in Well Table window.
⇒
Unselect wells
Cells can not be recognized.
It might be that the threshold of
binarization is too high.
⇒
First, reconfirm the intensity range
of image by using
“Image Viewer
function image acquisition
”
function, and, after that, reset the
threshold properly (and, set
“Object Detectivity” higher).
It might be that the range of
“Size
Filter
” is not porper.
⇒
Reset the range of
“Size Filter”
properly with visualization of
“Chart” function.
It might be that viewer status for is
turned to undisplayed (
).
⇒
Turn it to displayed status (
).
(Refer to page
It might be that Area window is
selected in ImageViewer
⇒
Recognized result is not shown in
Area window. Select Image View
window. (refer to page
It might be that Z position that sample
doesn’t exist is shown by viewer.
⇒
Adjust Z position.
The defined line of boundary on cells
is not smooth.
It might be that the intensity on the
boundary fluctuates or is too dark.
⇒
Increase the intensity of image by
making the exposure time longer
or increasing laser output power.
⇒
To define outer envelope of the
boundary as the resultant
boundary, add
“Closing Circle”
command to
“Binarize Transform”
block by using
“Edit” of “Analysis”
window.
⇒
To define inner envelope of the
boundary as the resultant
boundary, add
“Opening Circle”
command to
“Binarize Transform”
block by using
“Edit” of “Analysis”
window.
How to clear recognition showing?
⇒
Turn viewer status to undisplayed
status (
Содержание CQ1
Страница 2: ...This document corresponds to the following versions of software Measurement software R1 04 ...
Страница 8: ...vi IM 80J01A01 01E Right Side of Utility Box Aperture label ...
Страница 24: ......
Страница 34: ......
Страница 48: ......
Страница 177: ... 5 Functions of CQ1 Software 5 81 IM 80J01A01 01E Drag tab to move the window ...
Страница 178: ...5 82 5 Functions of CQ1 Software IM 80J01A01 01E Drag tab to square mark as following to add the tab to the window ...
Страница 187: ... 5 Functions of CQ1 Software 5 91 IM 80J01A01 01E 4 Area Window Display image of Area by tile Change to Area from Field ...
Страница 236: ......
Страница 238: ......
Страница 252: ...7 14 7 Trouble Shooting IM 80J01A01 01E Set sub cover and tighten 2 screws ...
Страница 256: ......
Страница 266: ......