<6. Parameter Lists>
213
IM 01E21A02-03EN
Relative
Index
Index
Parameter
Name
Default Value
Write
Mode
Description
AI1 AI2 AI3
AI4
(only
AXG)
29
4029 4129 4229 4329 LO_PRI
0
AUTO
Specifies the priority for the LO
alarm. Meanwhile, to activate
an alarm on communication,
it is necessary to specify the
priority for alarm activation in
advance. To activate an alarm,
set 3 or more. By sending with
priority information being added
in a communication frame
when the alarm is activated, for
example, the alarm is used to
apply a filter to ignore values
below the specified priority on
the host side.
30
4030 4130 4230 4330 LO_LIM
-FLT_MAX
AUTO
Specifies a limit value for the
LO alarm.
31
4031 4131 4231 4331 LO_LO_PRI
0
AUTO
Specifies the priority of the
LO_LO alarm. Meanwhile,
to activate an alarm on
communication, it is necessary
to specify the priority for
alarm activation in advance.
To activate an alarm, set 3 or
more. By sending with priority
information being added in a
communication frame when the
alarm is activated, for example,
the alarm is used to apply a
filter to ignore values below the
specified priority on the host
side.
32
4032 4132 4232 4332 LO_LO_LIM
-FLT_MAX
AUTO
Specifies a limit value for the
LO_LO alarm.
33
4033 4133 4233 4333 HI_HI_ALM
0(Uninitialized)
0(Uninitialized)
0,0
0(Other)
0.0
AUTO
-
-
-
-
Indicates information about the
HI_HI alarm that occurred.
34
4034 4134 4234 4334 HI_ALM
0(Uninitialized)
0(Uninitialized)
0,0
0(Other)
0.0
AUTO
-
-
-
-
Indicates information about the
HI alarm that occurred.
35
4035 4135 4235 4335 LO_ALM
0(Uninitialized)
0(Uninitialized)
0,0
0(Other)
0.0
AUTO
-
-
-
-
Indicates information about the
LO alarm that occurred.
36
4036 4136 4236 4336 LO_LO_ALM
0(Uninitialized)
0(Uninitialized)
0,0
0(Other)
0.0
AUTO
-
-
-
-
Indicates information about the
LO_LO alarm that occurred.
37
4037 4137 4237 4337
BLOCK_ERR_
DESC_1
0x00000000
-
Bit assignment complies with
the Communication standard
specifications.