<5. Functions>
65
IM 01E21A02-03EN
F0407.ai
Autozero function
executing...
00 : 00
00 : 08
SET
SFT
INC
The time remaining
until the end.
The zero adjustment starts, and the progress is displayed with the remaining time and a
bar graph. Wait for the zero adjustment to complete.
F0408.ai
Autozero
Execute
Result
▲
▼
00 : 00
SET
SFT
INC
►
After the zero adjustment is finished, the display returns to “Autozero” menu.
When executing zero adjustment from F
OUNDATION
Fieldbus, zero adjustment is set with the
procedures of the interactive operation guide called DD Method. Follow procedures on the
operation screen.
Confirmation of zero adjustment result
The zero adjustment result using the display can be confirmed with the following parameter.
Menu path
Display
Device setup ► Diag/Service ► Autozero ► Result ► (see below)
F
OUNDATION
Fieldbus
Device Configuration ► STB ► Device Configuration ► Maintenance ► Autozero ► (see
below)
Block
Name
Relative
Index
Parameter
Description
Display
F
OUNDATION
Fieldbus
STB
60
Zero value
Zero Value
Refer to the following
confirmation methods for
results of zero adjustment and
the display.
F0409.ai
Result
Zero value
▲
▼
00 : 00
SET
SFT
INC
For the result of the zero adjustment, select “Result” and then “Zero value”.
F0410.ai
Zero value
– 00.018cm/s
– 00.018
cm/s
– 99.999
+ 99.999
00 : 00
SET
SFT
INC
The result of zero adjustment is displayed as on the left of the screen.
NOTE
When the zero adjustment result exceeds the defined value, the warning [092:AZ warn] is
displayed.