Using the OSICS ECL Module
126
OSICS User Manual
Once both calibration points are entered, the new calibration values are implemented
immediately and stored in the system's flash-memory, and the current wavelength/
frequency setting of the ECL module is updated to match the new calibration data.
The new calibration wavelengths/frequencies sets the new system calibration default.
The ECL
Unit Setup
menu is displayed.
9.9.3
Performing a Power Calibration
Subject
This section explains how to perform a two-point power calibration sequence using the
Power calibration
function. The purpose of this calibration is to correct the discrepancies
between the power displayed by the OSICS ECL module and the power measured at your
reference power meter.
Also, calibration is made at two different wavelengths to account for small wavelength-
dependence of the power meter.
The new calibration data sets the new system calibration default.
Before Starting
Make sure you have a power meter.
Procedure
1.
Set the ECL output power to
1 mW
.
2.
Connect the optical output of the OSICS ECL module to a power meter.
3.
Enable the module optical output (see section
Enabling/Disabling the Light Source of
4.
Access the ECL
Unit Setup
Accessing an OSICS Module Setup
).
5.
Expand the
Calibration
menu and enter the
Power calibration
sub-menu.
The message
Moving...
appears, and the
Power calibration
sub-menu appears,
prompting you to set the power at the first wavelength value.
Figure 90: ECL Unit Setup – Power calibration, power value at first wavelength
6.
Measure the power level at the first wavelength on the power meter.
A valid power value can range from 0.6 mW to 1.2 mW.
7.
Enter the power value measured by the power meter and select
ENTER
.
The message
Moving...
appears while the power value is set. The
Power
calibration
window appears, showing the second window. The ECL tunable laser
Main
EXIT
2
3
4
5
6
7
8
ECL Unit Setup - Power calibration
ESCAPE
First power value :
Correct the value.
ENTER
1
▲
P = +09.160 dBm
First wavelength : 1540.000 nm.
Содержание OSICS
Страница 2: ......
Страница 6: ...About This Manual 6 OSICS User Manual...
Страница 28: ...Installing and Connecting the OSICS Multifunction Platform 28 OSICS User Manual...
Страница 56: ...Using the OSICS ATN HP Module 56 OSICS User Manual BKR TLS 50 ATN DFB ECL T100 SLD SWT TLS AG SWT APC MAIN...
Страница 64: ...Using the OSICS BKR Module 64 OSICS User Manual BKR TLS 50 ATN DFB ECL T100 SLD SWT TLS AG SWT APC MAIN...
Страница 98: ...Using the OSICS DFB Module 98 OSICS User Manual BKR TLS 50 ATN DFB ECL T100 SLD SWT TLS AG SWT APC MAIN...
Страница 136: ...Using the OSICS SLD Module 136 OSICS User Manual BKR TLS 50 ATN DFB ECL T100 SLD SWT TLS AG SWT APC MAIN...
Страница 248: ...Certification and Compliance 248 OSICS User Manual...