Controlling the Optical Output of all OSICS Modules Simultaneously
38
OSICS User Manual
4.2
Configuring the Optical Output Power of All Modules
4.2.1
Selecting the Power Unit for All Installed Modules
Subject
The following procedure explains how to modify the power unit (dBm or mW) for all
modules installed in the OSICS Mainframe. You cannot set the power unit of an individual
module.
Procedure
1.
Access the
Main Setup
menu (see section
Accessing the OSICS Mainframe Main Setup
).
2.
Expand the
Power
menu and enter the
Set unit dBm/mW
sub-menu.
The
Set unit dBm/mW
sub-menu appears and displays the current unit setting.
Figure 12: Main Setup – Set unit dBm/mW sub-menu
3.
Turn the rotary knob to put the cursor before the wanted unit and press it to validate
your selection.
The
Main Setup
menu is automatically displayed and the optical power value of each
module is now set to the selected unit.
4.2.2
Setting the Optical Output Power of All Installed Modules
Subject
You can set the optical output power in two different ways:
•
From the wanted module menu, you can set the module output power individually.
For more details on this procedure, see section
Setting the Optical Output Power of
the Module
in the appropriate module section of this manual.
•
From the OSICS Mainframe
Main Setup
menu, you can set all modules output power
to the same value simultaneously, as detailed in the following procedure.
As modules of different types and different power specifications may be installed in
the system, the power setting may exceed one of the modules maximum power-
rating. To prevent any possible power-overshoots, the power of each module is
software-limited within the acceptable range for that particular module.
Procedure
1.
Access the
Main Setup
menu (see section
Accessing the OSICS Mainframe Main Setup
).
2.
Expand the
Power
menu and enter the
Set power reference
sub-menu.
The
Power reference
sub-menu appears and displays the current power setting.
Main
EXIT
1
2
3
4
5
6
7
8
Main Setup - Set unit dBm/mW
ESCAPE
▲
Power Unit :
dBm
mW
Select a new unit.
Содержание OSICS
Страница 2: ......
Страница 6: ...About This Manual 6 OSICS User Manual...
Страница 28: ...Installing and Connecting the OSICS Multifunction Platform 28 OSICS User Manual...
Страница 56: ...Using the OSICS ATN HP Module 56 OSICS User Manual BKR TLS 50 ATN DFB ECL T100 SLD SWT TLS AG SWT APC MAIN...
Страница 64: ...Using the OSICS BKR Module 64 OSICS User Manual BKR TLS 50 ATN DFB ECL T100 SLD SWT TLS AG SWT APC MAIN...
Страница 98: ...Using the OSICS DFB Module 98 OSICS User Manual BKR TLS 50 ATN DFB ECL T100 SLD SWT TLS AG SWT APC MAIN...
Страница 136: ...Using the OSICS SLD Module 136 OSICS User Manual BKR TLS 50 ATN DFB ECL T100 SLD SWT TLS AG SWT APC MAIN...
Страница 248: ...Certification and Compliance 248 OSICS User Manual...