12
SAFETY DEVICES AND ERROR CODES
Purposes and Operations of Safety Device and Counter-measures
When an abnormal condition occurs, an error code is displayed in the main display. Immediate
action should be taken according to the specific counter-measures.
Er. No
Error name
Main Display
Sub display
Er.00
Communication Er
Communication Er
Ex
a
connection
Er.01
Sensor Error
Sensor Error
Calls a service
Er.02
SSR Error
SSR Error
Calls a service
Er.03
Heat Error
Heat Error
Calls a service
Er.04
Fan Error
Fan Error
Calls a service
Er.05
Auto-damper Er
Auto-damper Er
Calls a service
Er.07
Overheat Error
Overheat Error
Calls a service
Er.08
Controller Er
Controller Er
Calls a service
Er.10
Main relay Er
Main relay Er
Calls a service
Er.14
RAM Error
Controller Er
Calls a service
Er.15
EEPROM Error
Controller Er
Calls a service
How to manage
●
In case these error code mentioned above is displayed, recode the error code and shut
off the power immediately.
●
In case “Er14” is displayed, once turn off a circuit breaker. Then, after more than 30
seconds, turn on the breaker again. In case “Er14” is still displayed even after the
breaker is turned on, then contact a customer support center.
●
In case any abnormal condition occurs, parts replacement or device inspection should
be taken. Contact a customer support center. In contacting, never fail to inform of
error number. Refer to P 57 (AFTER SERVICE AND WARRANTY).