33
PROGRAMMED OPERATION METHOD
TO EDIT SEGMENT
・
There are 12 kinds of setup items of 1 segment.
(
Full option installation
)
No.
Name
Setup range
Option
1
Ramp time
Step
、
1 min
~
999 hr, End
2
Setup temp
A setup temperature range by the model
3
Soak time
0
m
in
~
999 hr, Hold
4
Wait mode
OFF, ON
5
Fan
OFF, 1
~
10
(
or ON
)※
○
6
Damper
0%, 25%, 50%, 75%, 100%
○
7
Event 1
OFF, ON
○
8
Event 2
OFF, ON
○
9
Event 3
OFF, ON
○
10
Event 4
OFF, ON
○
11
Rep. start segment
No, Segment number that 1 - 16 have been
registered
12
Rep count
1
~
9999, Infinity
※
In case “Wind velocity variable device” option is not equipped, you can choose “On” or
“Off” of fan only by “End segment”
・
Concept of segment is shown as below. However “Start temp” corresponds to
“Temperature of the inside of chamber” when you carry out segment number 1 just after
program starts. Other than that case it will be “Setup temp” of segment, which is carried
out most recently.
Ramp time
Sork time
Time
Setup temp
Temp
Start temp
・
Concept of repeat is shown as below. The number of repeat times does not include the
first practice of repeat section.
Segment No.4
Repeat start
segment :2
Repeat time :5
Segment No.1
There is no
order for a
repeat.
Segment No.2
There is no
order for a
repeat.
Segment No.3
There is no
order for a
repeat.
Segment No.5
There is no
order for a
repeat.
Repeat section
Repeat×5