47
SETTING THE CALIBRATION OFFSET FUNCTION
1.
To choose “Calibration offset”
・
On item choice screen in function menu, have “Calibrate” display and press jog dial.
Calibrate >
±
0
℃
℃
FIXED TEMP
AUTO START
AUTO STOP
PROGRAM
2.
Input of “Offset temperature”
・
“►” cursor disappears and current offset temperature blinks. Turn jog dial to set tem-
perature and have it display.
Calibrate
±0℃
↓
Calibrate
+5℃
3.
Press “Jog dial”
Calibrate >
+
5
℃
℃
FIXED TEMP
AUTO START
AUTO STOP
PROGRAM
・
Calibration offset temperature is fixed and it returns to item choice screen in function
menu.